Taxa de sucesso de posicionamento de cabo-eletrodo ventricular esquerdo em região-alvo sem utilização de bainha e/ou venografia durante procedimento de implante de marcapasso biventricular / Success rate of left ventricular lead positioning in target region without a sheath and/or venography during biventricular pacemaker implantation procedure
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm
; 30(2): f:43-l:46, abr.-jun. 2017. ilus, graf
Article
in Portuguese
| LILACS
| ID: biblio-848020
Responsible library:
BR44.1
Localization: BR44.1
RESUMO
Introdução:
Este trabalho teve por objetivo avaliar a taxa de sucesso de implante de cabo-eletrodo ventricular em região lateral sem uso de venografia.Método:
Foram analisados 100 casos de implante de cabo-eletrodo ventricular esquerdo ao longo de 30 meses, em um único centro, quanto a sua localização e calculada a taxa de implantes em parede lateral (ântero-lateral, lateral e póstero-lateral). Todos os procedimentos foram realizados sem uso de bainha e sem realização de venografia, por meio de técnica de cateterização do seio coronário baseada no componente atrial do eletrograma endocavitário.Resultados:
Em 83% dos casos foi conseguido implante em parede lateral, predominantemente em paredes póstero-lateral e lateral, com tempo de radioscopia médio de 5,97 minutos. Em 10% houve insucesso, com necessidade de implante de cabo-eletrodo epicárdico.Conclusão:
O implante de cabo-eletrodo ventricular esquerdo em região lateral sem uso de venografia baseado no componente atrial do eletrograma endocavitário constitui técnica segura e eficaz, apresentando ainda redução do tempo de radioscopiaABSTRACT
Background:
This study was aimed at evaluating left ventricular pacing implant success rate in the lateral region without the use of venography.Method:
We analyzed 100 cases of left ventricular lead implants over 30 months in a single center as to the location of the left ventricular lead and calculated the rate of implants in the lateral wall (anterolateral, lateral and posterolateral). All procedures were carried out without the use of a sheath and without performing venography using the coronary sinus catheter technique based on the atrial endocardial electrogram component.Results:
In 83% of cases it was successfully implanted in the lateral wall, predominantly in the posterolateral and lateral walls, with a mean radioscopy time of 5.97 minutes. It failed in 10% of the cases and we had to use an epicardial lead.Conclusion:
Left ventricular lead implantation in the lateral region without venography based on atrial component of the endocavitary electrogram is a safe and effective technique, which also reduces radioscopy time
Collection:
International databases
Database:
LILACS
Main subject:
Pacemaker, Artificial
/
Prostheses and Implants
/
Phlebography
/
Electrodes
/
Heart Ventricles
Type of study:
Diagnostic study
/
Etiology study
/
Incidence study
/
Observational study
/
Risk factors
Limits:
Humans
Language:
Portuguese
Journal:
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm
Journal subject:
Cardiology
Year:
2017
Document type:
Article
Affiliation country:
Brazil
Institution/Affiliation country:
Hospital Beneficência Portuguesa de São Paulo/BR