Your browser doesn't support javascript.
loading
Avaliação da radiação espalhada e do impacto dos dispositivos locais de proteção em laboratório de cardiologia intervencionista / Evaluation of scattered radiation and impact of local protective devices in an interventional cardiology laboratory
Bienert, Igor Ribeiro de Castro; Ferreira Junior, Luiz Carlos; Silva, Paulo André da; Miwa, Daniela Tomie Kasama; Florêncio, Carla Liberato Bastos; Mota, Rodrigo Lupp; Rinaldi, Fabio Salermo; Andrade, Pedro Beraldo de.
Afiliação
  • Bienert, Igor Ribeiro de Castro; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Ferreira Junior, Luiz Carlos; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Silva, Paulo André da; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Miwa, Daniela Tomie Kasama; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Florêncio, Carla Liberato Bastos; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Mota, Rodrigo Lupp; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Rinaldi, Fabio Salermo; Faculdade Estadual de Medicina de Marília. Hospital das Clínicas. Marília - SP. BR
  • Andrade, Pedro Beraldo de; Santa Casa de Misericórdia de Marília. Marília - SP. BR
Rev. bras. cardiol. invasiva ; 24(1-4): 38-43, jan.-dez. 2016. tab, graf
Artigo em Português | LILACS | ID: biblio-878999
Biblioteca responsável: BR44.1
Localização: BR44.1
RESUMO

Introdução:

As aplicações médicas representam a maior fonte de exposição radiológica ionizante e, neste contexto, é de especial importância a questão do risco profissional. Embora existam diferentes sensibilidades à radiação dos distintos órgãos do corpo humano, avaliações específicas do impacto nas diversas regiões do corpo do operador, com diferentes dispositivos de radioproteção, são raras em nosso meio.

Métodos:

Teste de radiação espalhada foi realizado com câmara de ionização em estação de fluoroscopia, com jogo de acessórios de radioproteção padrão do equipamento (saia inferior e escudo móvel superior, em duas diferentes posições), a distâncias sequenciais em relação à fonte, utilizando fantoma de acrílico em simulação de tórax humano.

Resultados:

Foram identificadas diferenças na radiação em relação à distância e ao uso dos dispositivos de radioproteção. A redução mediana da radiação foi de 50,6% (intervalo interquartil − IQ de 39,42% a 51,05%) com uso do escudo saia inferior, 71,3% (IQ de 67,66% a 77,05%) com adição de escudo superior em posicionamento angulado e 84,7% (IQ de 83,75% a 85,87%) com adição de escudo superior em linha ao escudo inferior. Diferenças significativas foram encontradas ainda em relação à altura e à distância da fonte.

Conclusões:

O uso dos dispositivos locais de radioproteção avaliados se mostrou efetivo na redução global do impacto radiológico ao operador, havendo, no entanto, vias de escape de radiação, especialmente com posicionamento não ideal, demonstrando a importância do uso adicional dos dispositivos de proteção individuais
ABSTRACT

Background:

Medical applications are the main source of ionizing radiation exposure, and in this context the issue of occupational risk is particularly important. Although the different organs of the human body present different radiation sensitivities, specific assessments of the impact on the different regions of the interventionist's body with diverse radioprotection devices are rare in Brazil.

Methods:

A scattered radiation test was performed using an ionization chamber in a fluoroscopy station, with standard radioprotection accessory kit of the equipment (lower skirt and upper movable shield, in two different positions), at sequential distances from the source, using acrylic phantoms as human chest simulation.

Results:

Differences in radiation were identified in relation to distance and use of radioprotection devices. The median radiation reduction was 50.6% (interquartile range ­ IQ from 39.42% to 51.05%) using the lower skirt shield, 71.3% (IQ from 67.66% to 77.05%) with the addition of an upper shield in angulated position, and 84.7% (IQ from 83.75% to 85.87%) with the addition of an upper shield aligned with the lower shield. Significant differences were also found regarding height and distance from the source

Conclusions:

The use of the assessed local radioprotection devices was effective in reducing the overall radiological impact to the interventionist. However, there were radiation escape routes, especially with non-ideal positioning, demonstrating the importance of the additional use of individual protection devices
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Radiação Ionizante / Proteção Radiológica / Diagnóstico por Imagem / Equipamento de Proteção Individual Tipo de estudo: Estudo diagnóstico / Estudo de etiologia / Guia de prática clínica / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Assunto da revista: Cardiologia / CIRURGIA GERAL Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculdade Estadual de Medicina de Marília/BR / Santa Casa de Misericórdia de Marília/BR

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Radiação Ionizante / Proteção Radiológica / Diagnóstico por Imagem / Equipamento de Proteção Individual Tipo de estudo: Estudo diagnóstico / Estudo de etiologia / Guia de prática clínica / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Assunto da revista: Cardiologia / CIRURGIA GERAL Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculdade Estadual de Medicina de Marília/BR / Santa Casa de Misericórdia de Marília/BR
...