Your browser doesn't support javascript.
loading
Portadores de CDI com limiar de desfibrilação elevado: evolução clínica e alternativas terapêuticas / ICD patients with elevated defibrillation threshold: clinical behavior and therapeutic alternatives
Lima, Carlos Eduardo Batista de; Martinelli Filho, Martino; Silva, Rodrigo Tavares; Tamaki, Wagner Tetsuji; Oliveira, Júlio Cesar de; Martins, Daniela Cabral; Nishióka, Silvana Angelina D'Orio; Pedrosa, Anísio Alexandre Andrade; Siqueira, Sérgio Freitas; Costa, Roberto.
Afiliação
  • Lima, Carlos Eduardo Batista de; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Martinelli Filho, Martino; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Silva, Rodrigo Tavares; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Tamaki, Wagner Tetsuji; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Oliveira, Júlio Cesar de; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Martins, Daniela Cabral; s.af
  • Nishióka, Silvana Angelina D'Orio; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Pedrosa, Anísio Alexandre Andrade; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Siqueira, Sérgio Freitas; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
  • Costa, Roberto; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. BR
Arq. bras. cardiol ; 90(3): 177-184, mar. 2008. ilus, graf, tab
Artigo em Inglês, Português | LILACS | ID: lil-479618
Biblioteca responsável: BR1.1
RESUMO
FUNDAMENTO A programação ideal da energia de choque do CDI deve ser pelo menos 10 J acima do limiar de desfibrilação (LDF), necessitando de técnicas alternativas quando o LDF é elevado.

OBJETIVO:

Avaliar o comportamento clínico dos portadores de CDI com LDF>25 J e a eficácia da terapêutica escolhida.

MÉTODOS:

Foram selecionados portadores de CDI, entre janeiro de 2000 e agosto de 2004 (banco de dados prospectivo), com LDF>25 J intra-operatório, e analisaram-se características clínicas, FEVE, resgate de eventos arrítmicos pelo CDI e óbitos.

RESULTADOS:

dentre 476 pacientes, 16 (3,36 por cento) apresentaram LDF>25J. Idade média de 56,5 anos, sendo 13 pacientes (81 por cento) do sexo masculino. Quanto à cardiopatia de base 09 eram chagásicos, 04 isquêmicos e 03 com etiologia idiopática. A FEVE média dos pacientes foi 37 por cento e 94 por cento utilizavam amiodarona. O seguimento médio foi de 25,3 meses. Em 02 pacientes com LDF > Choque Máximo (CM), foi necessário implante de eletrodo de choque adicional (array), sendo mantido programação com CM em zona de FV (>182bpm) nos demais. Durante o seguimento 03 pacientes apresentaram 67 terapias de choque apropriadas (TCA) com sucesso. Ocorreram 07 óbitos sendo 5 por causas não cardíacas e 2 por insuficiência cardíaca avançada. Os pacientes que foram a óbito apresentaram níveis de LDF maiores (p=0,0446), entretanto sem relação com a causa dos mesmos tendo em vista que não ocorreram TCA sem sucesso.

CONCLUSÃO:

Nessa coorte de pacientes com CDI, a ocorrência de LDF elevado foi baixa, implicando terapêuticas alternativas. Houve associação com disfunção ventricular grave, entretanto sem correlação com as causas de óbito.
ABSTRACT

BACKGROUND:

The ideal programming of the implantable cardioverter defibrillator (ICD) shock energy should be at least 10J above the defibrillation threshold (DFT), requiring alternative techniques when the DFT is elevated.

OBJECTIVE:

To assess the clinical behavior of ICD patients with DFT>25J and the efficacy of the chosen therapy.

METHODS:

Patients who had undergone ICD implantation between Jan/00 and Aug/04 (prospective database) and presented intraoperative DFT>25J were selected. The analyzed variables were clinical characteristics, LVEF, rescue of arrhythmic events from ICD and causes of deaths.

RESULTS:

among 476 patients, 16 (3.36 percent) presented DFT>25J. The mean age was 56.5 years, and 13 patients (81 percent) were men. According to the baseline cardiomyopathy, 09 patients had Chagas' disease, 04 had ischemic cardiomyopathy and 03 had idiopathic cardiomyopathy. Mean LVEF was 0.37 and amiodarone was used by 94 percent of the patients. Mean follow-up (FU) period was 25.3 months. DFT was higher than maximum energy shock (MES) in 2 patients and it was necessary to implant an additional shock electrode (array). It was programmed MES in ventricular fibrillation zone of ICD therapy in the other patients. In the FU, 03 patients had 67 successful appropriate shock therapies (AST). There were 05 noncardiac and 02 heart failure deaths. The patients who died showed higher DFT levels (p=0.044) without correlation with death because there wasn't unsuccessful AST.

CONCLUSION:

In this cohort of ICD patients, the occurrence of elevated DFT (>25J) was low, leading to alternative therapies. There was an association with severe ventricular dysfunction, although without correlation to the causes of death.
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Contexto em Saúde: ODS3 - Saúde e Bem-Estar / Doenças Negligenciadas / ODS3 - Meta 3.3 Acabar com as doenças tropicais negligenciadas e combater as doenças transmissíveis / ODS3 - Meta 3.4 Reduzir as mortes prematuras devido doenças não transmissíveis Problema de saúde: Meta 3.4: Reduzir as mortes prematuras devido doenças não transmissíveis / Doença de Chagas / Doença de Chagas / Doença Cardiovascular / Outras Doenças Circulatórias Base de dados: LILACS Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis Tipo de estudo: Estudo de etiologia / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Inglês / Português Revista: Arq. bras. cardiol Assunto da revista: Cardiologia Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade de São Paulo/BR
Texto completo: Disponível Coleções: Bases de dados internacionais Contexto em Saúde: ODS3 - Saúde e Bem-Estar / Doenças Negligenciadas / ODS3 - Meta 3.3 Acabar com as doenças tropicais negligenciadas e combater as doenças transmissíveis / ODS3 - Meta 3.4 Reduzir as mortes prematuras devido doenças não transmissíveis Problema de saúde: Meta 3.4: Reduzir as mortes prematuras devido doenças não transmissíveis / Doença de Chagas / Doença de Chagas / Doença Cardiovascular / Outras Doenças Circulatórias Base de dados: LILACS Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis Tipo de estudo: Estudo de etiologia / Estudo observacional / Estudo prognóstico / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Inglês / Português Revista: Arq. bras. cardiol Assunto da revista: Cardiologia Ano de publicação: 2008 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade de São Paulo/BR
...