HPLC-APCI-MS for the determination of vitamin K(1) in human plasma: method and clinical application.

J Chromatogr B Analyt Technol Biomed Life Sci; 875(2): 541-5, 2008 Nov 15.
Artigo em Inglês | MEDLINE | ID: mdl-18952509
A sensitive HPLC-APCI-MS method for the determination of vitamin K(1) (VK-1) in human plasma was established. Target ions at [M+H](+)m/z 451.5 for VK-1 and [M+H](+)m/z 331.4 for the I.S. (teprenone). Calibration curve was linear over the range of 0.3-1,000 ng/ml. The lower limit of quantification was 0.3 ng/ml. The intra- and inter-batch variability values were less than 8% and 15%, respectively. The C(max) was 210.1+/-86.7 ng/ml while the elimination half-life (t(1/2)) was 8.8+/-1.7h and time to the C(max) was 5.5+/-0.8h after administration of soft capsule containing 10mg VK-1.