1.
Opt Lett ; 48(10): 2615-2618, 2023 May 15.
Artigo em Inglês | MEDLINE | ID: mdl-37186722

RESUMO

Fiber-optic hydrophones (FOHs) are widely used to detect high-intensity focused ultrasound (HIFU) fields. The most common type consists of an uncoated single-mode fiber with a perpendicularly cleaved end face. The main disadvantage of these hydrophones is their low signal-to-noise ratio (SNR). To increase the SNR, signal averaging is performed, but the associated increased acquisition times hinder ultrasound field scans. In this study, with a view to increasing SNR while withstanding HIFU pressures, the bare FOH paradigm is extended to include a partially reflective coating on the fiber end face. Here, a numerical model based on the general transfer-matrix method was implemented. Based on the simulation results, a single-layer, 172 nm TiO2-coated FOH was fabricated. The frequency range of the hydrophone was verified from 1 to 30 MHz. The SNR of the acoustic measurement with the coated sensor was 21 dB higher than that of the uncoated one. The coated sensor successfully withstood a peak positive pressure of 35 MPa for 6000 pulses.

2.
J Acoust Soc Am ; 149(4): 2732, 2021 04.
Artigo em Inglês | MEDLINE | ID: mdl-33940866

RESUMO

Laser-generated focused ultrasound (LGFU) transducers used for ultrasound therapy commonly have large diameters (6-15 mm), but smaller lateral dimensions (<4 mm) are required for interventional applications. To address the question of whether miniaturized LGFU transducers could generate sufficient pressure at the focus to enable therapeutic effects, a modelling and measurement study is performed. Measurements are carried out for both linear and nonlinear propagation for various illumination schemes and compared with the model. The model comprises several innovations. First, the model allows for radially varying acoustic input distributions on the surface of the LGFU transducer, which arise from the excitation light impinging on the curved transducer surfaces. This realistic representation of the source prevents the overestimation of the achievable pressures (shown here to be as high as 1.8 times). Second, an alternative inverse Gaussian illumination paradigm is proposed to achieve higher pressures; a 35% increase is observed in the measurements. Simulations show that LGFU transducers as small as 3.5 mm could generate sufficient peak negative pressures at the focus to exceed the cavitation threshold in water and blood. Transducers of this scale could be integrated with interventional devices, thereby opening new opportunities for therapeutic applications from inside the body.


Assuntos
Transdutores , Terapia por Ultrassom , Acústica , Lasers
3.
Comput Biol Med ; 109: 333-341, 2019 06.
Artigo em Inglês | MEDLINE | ID: mdl-31129310

RESUMO

Photoacoustic microscopy (PAM) is classified as a hybrid imaging technique based on the photoacoustic effect and has been frequently studied in recent years. Photoacoustic (PA) signals are inherently recorded in a noisy environment and are also exposed to noise by system components. Therefore, it is essential to reduce the noise in PA signals to reconstruct images with less error. In this study, an image reconstruction algorithm for PAM system was implemented and different filtering approaches for denoising were compared. Studies were carried out in three steps: simulation, experimental phantom and blood cell studies. FIR low-pass and band-pass filters and Discrete Wavelet Transform (DWT) based filters (mother wavelets: "bior3.5″, "bior3.7″, "sym7″) with four different thresholding techniques were examined. For the evaluation purposes, Root Mean Square Error (RMSE), Signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR) metrics were calculated. In the simulation studies, the most effective methods were obtained as: sym7/heursure/hard thresh. combination (low and medium level noise) and bior3.7/sqtwolog/soft thresh. combination (high-level noise). In experimental phantom studies, noise was classified into five levels. Different filtering approaches perform better depending on the SNR of PA images. For the blood cell study, based on the standard deviation in the background, sym7/sqtwolog/soft thresh. combination provided the best improvement and this result supported the experimental phantom results.


Assuntos
Processamento de Imagem Assistida por Computador , Microscopia , Modelos Teóricos , Técnicas Fotoacústicas , Razão Sinal-Ruído
4.
Sci Rep ; 6: 38674, 2016 12 08.
Artigo em Inglês | MEDLINE | ID: mdl-27929049

RESUMO

Photoacoustic imaging is based on the detection of generated acoustic waves through thermal expansion of tissue illuminated by short laser pulses. Fiber lasers as an excitation source for photoacoustic imaging have recently been preferred for their high repetition frequencies. Here, we report a unique fiber laser developed specifically for multiwavelength photoacoustic microscopy system. The laser is custom-made for maximum flexibility in adjustment of its parameters; pulse duration (5-10 ns), pulse energy (up to 10 µJ) and repetition frequency (up to 1 MHz) independently from each other and covers a broad spectral region from 450 to 1100 nm and also can emit wavelengths of 532, 355, and 266 nm. The laser system consists of a master oscillator power amplifier, seeding two stages; supercontinuum and harmonic generation units. The laser is outstanding since the oscillator, amplifier and supercontinuum generation parts are all-fiber integrated with custom-developed electronics and software. To demonstrate the feasibility of the system, the images of several elements of standardized resolution test chart are acquired at multiple wavelengths. The lateral resolution of optical resolution photoacoustic microscopy system is determined as 2.68 µm. The developed system may pave the way for spectroscopic photoacoustic microscopy applications via widely tunable fiber laser technologies.

5.
Artigo em Inglês | MEDLINE | ID: mdl-25215814

RESUMO

We investigate the radiation force on a microbubble due to the photoacoustic wave which is generated by using a pulsed laser. In particular, we focus on the dependence of pulsed laser parameters on the radiation force. In order to do so, we first obtain a new and comprehensive analytical solution to the photoacoustic wave equation based on the Fourier transform for various absorption profiles. Then, we write an expression of the radiation force containing explicit laser parameters, pulse duration, and beamwidth of the laser. Furthermore, we calculate the primary radiation force acting on a microbubble. We show that laser parameters and the position of the microbubble relative to a photoacoustic source have a considerable effect on the primary radiation force. By means of recent developments in laser technologies that render tunability of pulse duration and repetition frequency possible, an adjustable radiation force can be applied to microbubbles. High spatial control of applied force is ensured on account of smaller focal spots achievable by focused optics. In this context, conventional piezoelectric acoustic source applications could be surpassed. In addition, it is possible to increase the radiation force by making source wavelength with the absorption peak of absorber concurrent. The application of photoacoustic radiation force can open a cache of opportunities such as manipulation of microbubbles used as contrast agents and as carrier vehicles for drugs and genes with a desired force along with in vivo applications.


Assuntos
Lasers , Microbolhas , Modelos Teóricos , Radiação , Absorção de Radiação , Acústica , Análise de Fourier , Óptica e Fotônica