Electrophoresis ; 2024 Apr 30.
Artigo em Inglês | MEDLINE | ID: mdl-38687164

RESUMO

Capillary temperature control during capillary electrophoresis (CE) separations is key for achieving accurate and reproducible results with a broad array of potential methods. However, the difficulty of enabling typical fluid temperature control loops on portable instruments has meant that active capillary temperature control of in situ CE systems has frequently been overlooked. This work describes construction and test of a solid-state device for capillary temperature control that is suitable for inclusion with in situ instruments, including those designed for space missions. Two test articles were built, a thermal mass model (TMM) and a functional model (FM). The TMM demonstrated that temperature gradients could be limited using the proposed control scheme, and that our thermal modeling of the system can be relied on for future adaptations of physical geometries of the system. The FM demonstrated CE analytical performance while under active temperature control and that the device was compatible with the harsh thermal-vacuum environments that might be encountered during space flight.