Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 8 de 8
Filtrar
Mais filtros

Base de dados
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
Biomed Microdevices ; 25(4): 41, 2023 10 23.
Artigo em Inglês | MEDLINE | ID: mdl-37870619

RESUMO

Reliability evaluation results of a manufacturable 32-channel cochlear electrode array are reported in this paper. Applying automated laser micro-machining process and a layer-by-layer silicone deposition scheme, authors developed the manufacturing methods of the electrode array for fine patterning and mass production. The developed electrode array has been verified through the requirements specified by the ISO Standard 14708-7. And the insertion trauma of the electrode array has been evaluated based on human temporal bone studies. According to the specified requirements, the electrode array was assessed through elongation & insulation, flexural, and fatigue tests. In addition, Temporal bone study was performed using eight fresh-frozen cadaver temporal bones with the electrode arrays inserted via the round window. Following soaking in saline condition, the impedances between conducting wires of the electrode array were measured over 100 kΩ (the pass/fail criterion). After each required test, it was shown that the electrode array maintained the electrical continuity and insulation condition. The average insertion angle of the electrode array inside the scala tympani was 399.7°. The human temporal bone studies exhibited atraumatic insertion rate of 60.3% (grade 0 or 1). The reliability of the manufacturable electrode array is successfully verified in mechanical, electrical, and histological aspects. Following the completion of a 32-channel cochlear implant system, the performance and stability of the 32-channel electrode array will be evaluated in clinical trials.


Assuntos
Implante Coclear , Implantes Cocleares , Humanos , Implante Coclear/métodos , Reprodutibilidade dos Testes , Rampa do Tímpano/cirurgia , Janela da Cóclea , Osso Temporal/cirurgia , Cóclea/cirurgia , Eletrodos Implantados
2.
J Neurosci ; 36(19): 5314-27, 2016 05 11.
Artigo em Inglês | MEDLINE | ID: mdl-27170128

RESUMO

UNLABELLED: Acetylcholine and noradrenaline are major neuromodulators that affect sensory processing in the cortex. Modality-specific sensory information is processed in defined areas of the cortex, but it is unclear whether cholinergic neurons in the basal forebrain (BF) and noradrenergic neurons in the locus ceruleus (LC) project to and modulate these areas in a sensory modality-selective manner. Here, we mapped BF and LC projections to different sensory cortices of the mouse using dual retrograde tracing. We found that while the innervation of cholinergic neurons into sensory cortices is predominantly modality specific, the projections of noradrenergic neurons diverge onto multiple sensory cortices. Consistent with this anatomy, optogenetic activation of cholinergic neurons in BF subnuclei induces modality-selective desynchronization in specific sensory cortices, whereas activation of noradrenergic LC neurons induces broad desynchronization throughout multiple sensory cortices. Thus, we demonstrate a clear distinction in the organization and function of cholinergic BF and noradrenergic LC projections into primary sensory cortices: cholinergic BF neurons are highly selective in their projections and modulation of specific sensory cortices, whereas noradrenergic LC neurons broadly innervate and modulate multiple sensory cortices. SIGNIFICANCE STATEMENT: Neuromodulatory inputs from the basal forebrain (BF) and locus ceruleus (LC) are widespread in the mammalian cerebral cortex and are known to play important roles in attention and arousal, but little is known about the selectivity of their cortical projections. Using a dual retrobead tracing technique along with optogenetic stimulation, we have identified anatomic and functional differences in the way cholinergic BF neurons and noradrenergic LC neurons project into primary sensory cortices. While BF projections are highly selective to individual sensory cortices, LC projections diverge into multiple sensory cortices. To our knowledge, this is the first definitive proof that BF and LC projections to primary sensory cortices show both anatomic and functional differences in selectivity for modulating cortical activity.


Assuntos
Locus Cerúleo/fisiologia , Prosencéfalo/fisiologia , Córtex Sensório-Motor/fisiologia , Transmissão Sináptica , Neurônios Adrenérgicos/citologia , Neurônios Adrenérgicos/fisiologia , Animais , Neurônios Colinérgicos/citologia , Neurônios Colinérgicos/fisiologia , Feminino , Locus Cerúleo/citologia , Masculino , Camundongos , Prosencéfalo/citologia , Córtex Sensório-Motor/citologia
3.
Micromachines (Basel) ; 12(7)2021 Jun 30.
Artigo em Inglês | MEDLINE | ID: mdl-34209329

RESUMO

(1) Background: In this study, we introduce a manufacturable 32-channel cochlear electrode array. In contrast to conventional cochlear electrode arrays manufactured by manual processes that consist of electrode-wire welding, the placement of each electrode, and silicone molding over wired structures, the proposed cochlear electrode array is manufactured by semi-automated laser micro-structuring and a mass-produced layer-by-layer silicone deposition scheme similar to the semiconductor fabrication process. (2) Methods: The proposed 32-channel electrode array has 32 electrode contacts with a length of 24 mm and 0.75 mm spacing between contacts. The width of the electrode array is 0.45 mm at its apex and 0.8 mm at its base, and it has a three-layered arrangement consisting of a 32-channel electrode layer and two 16-lead wire layers. To assess its feasibility, we conducted an electrochemical evaluation, stiffness measurements, and insertion force measurements. (3) Results: The electrochemical impedance and charge storage capacity are 3.11 ± 0.89 kOhm at 1 kHz and 5.09 mC/cm2, respectively. The V/H ratio, which indicates how large the vertical stiffness is compared to the horizontal stiffness, is 1.26. The insertion force is 17.4 mN at 8 mm from the round window, and the maximum extraction force is 61.4 mN. (4) Conclusions: The results of the preliminary feasibility assessment of the proposed 32-channel cochlear electrode array are presented. After further assessments are performed, a 32-channel cochlear implant system consisting of the proposed 32-channel electrode array, 32-channel neural stimulation and recording IC, titanium-based hermetic package, and sound processor with wireless power and signal transmission coil will be completed.

4.
Micromachines (Basel) ; 11(8)2020 Jul 24.
Artigo em Inglês | MEDLINE | ID: mdl-32722193

RESUMO

The authors would like to make the following changes to the published paper [...].

5.
Micromachines (Basel) ; 11(5)2020 May 25.
Artigo em Inglês | MEDLINE | ID: mdl-32466300

RESUMO

To investigate the neuronal visual encoding process in the retina, researchers have performed in vitro and ex vivo electrophysiological experiments using animal retinal tissues. The microelectrode array (MEA) has become a key component in retinal experiments because it enables simultaneous neural recording from a population of retinal neurons. However, in most retinal experiments, it is inevitable that the retinal tissue is flattened on the planar MEA, becoming deformed from the original hemispherical shape. During the tissue deforming process, the retina is subjected to mechanical stress, which can induce abnormal physiological conditions. To overcome this problem, in this study, we propose a hemispherical MEA with a curvature that allows retinal tissues to adhere closely to electrodes without tissue deformation. The electrode array is fabricated by stretching a thin, flexible polydimethylsiloxane (PDMS) electrode layer onto a hemispherical substrate. To form micro patterns of electrodes, laser processing is employed instead of conventional thin-film microfabrication processes. The feasibility for neural recording from retinal tissues using this array is shown by conducting ex vivo retinal experiments. We anticipate that the proposed techniques for hemispherical MEAs can be utilized not only for ex vivo retinal studies but also for various flexible electronics.

6.
Med Biol Eng Comput ; 58(2): 383-399, 2020 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-31853774

RESUMO

The osseous regeneration of large bone defects is still a major clinical challenge in maxillofacial and orthopedic surgery. Previous studies demonstrated that biphasic electrical stimulation (ES) stimulates bone formation; however, polyimide electrode should be removed after regeneration. This study presents an implantable electrical stimulation bioreactor with electrodes based on liquid crystal polymer (LCP), which can be permanently implanted due to excellent biocompatibility to bone tissue. The bioreactor was implanted into a critical-sized bone defect and subjected to ES for one week, where bone regeneration was evaluated four weeks after surgery using micro-CT. The effect of ES via the bioreactor was compared with a sham control group and a positive control group that received recombinant human bone morphogenetic protein (rhBMP)-2 (20 µg). New bone volume per tissue volume (BV/TV) in the ES and rhBMP-2 groups increased to 132% (p < 0.05) and 174% (p < 0.01), respectively, compared to that in the sham control group. In the histological evaluation, there was no inflammation within the bone defects and adjacent to LCP in all the groups. This study showed that the ES bioreactor with LCP electrodes could enhance bone regeneration at large bone defects, where LCP can act as a mechanically resistant outer box without inflammation. Graphical abstract To enhance bone regeneration, a bioreactor comprising collagen sponge and liquid crystal polymer-based electrode was implanted in the bone defect. Within the defect, electrical current pulses having biphasic waveform were applied from the implanted bioreactor.


Assuntos
Reatores Biológicos , Regeneração Óssea/fisiologia , Mandíbula/patologia , Mandíbula/fisiopatologia , Polímeros/química , Animais , Osso Esponjoso/diagnóstico por imagem , Osso Esponjoso/patologia , Osso Esponjoso/fisiopatologia , Estimulação Elétrica , Eletrodos , Masculino , Mandíbula/diagnóstico por imagem , Osteogênese , Coelhos , Microtomografia por Raio-X
7.
J Neurosci Methods ; 278: 25-35, 2017 02 15.
Artigo em Inglês | MEDLINE | ID: mdl-28040494

RESUMO

BACKGROUND: Ideally, neural probes should have channels with a three-dimensional (3-D) configuration to record the activities of 3-D neural circuits. Many types of 3-D neural probes have been developed; however, most of them were designed as an array of multiple shanks with electrodes located along one side of the shanks. NEW METHOD: We developed a novel liquid crystal polymer (LCP)-based neural probe with four-sided electrodes. This probe has electrodes on four sides of the shank, i.e., the front, back and two sidewalls. To generate the proposed configuration of the electrodes, we used a thermal lamination process involving LCP films and laser micromachining. RESULTS: The proposed novel four-sided neural probe, was used to successfully perform in vivo multichannel neural recording in the mouse primary somatosensory cortex. COMPARISON WITH EXISTING METHOD: The multichannel neural recording showed that the proposed four-sided neural probe can record spiking activities from a more diverse neuronal population than single-sided probes. This was confirmed by a pairwise Pearson correlation coefficient (Pearson's r) analysis and a cross-correlation analysis. CONCLUSION: The developed four-sided neural probe can be used to record various signals from a complex neural network.


Assuntos
Microeletrodos , Polímeros , Potenciais de Ação , Animais , Impedância Elétrica , Desenho de Equipamento , Camundongos Endogâmicos C57BL , Microscopia Eletrônica de Varredura , Microtecnologia , Neurônios/fisiologia , Córtex Somatossensorial/fisiologia
8.
Artigo em Inglês | MEDLINE | ID: mdl-24110931

RESUMO

Liquid Crystal Polymer (LCP) has been considered as an alternative biomaterial for implantable biomedical devices primarily for its low moisture absorption rate compared with conventional polymers such as polyimide, parylene and silicone elastomers. A novel retinal prosthetic device based on monolithic encapsulation of LCP is being developed in which entire neural stimulation circuitries are integrated into a thin and eye-conformable structure. Micromachining techniques for fabrication of a LCP retinal electrode array have been previously reported. In this research, however, for being used as a part of the LCP-based retinal implant, we developed advanced fabrication process of LCP retinal electrode through new approaches such as electroplating and laser-machining in order to achieve higher mechanical robustness, long-term reliability and flexibility. Thickened metal tracks could contribute to higher mechanical strength as well as higher long-term reliability when combined with laser-ablation process by allowing high-pressure lamination. Laser-thinning technique could improve the flexibility of LCP electrode.


Assuntos
Polímeros , Desenho de Prótese/métodos , Retina/fisiologia , Próteses Visuais , Impedância Elétrica , Eletrodos Implantados , Galvanoplastia , Lasers , Microeletrodos , Microscopia Eletrônica de Varredura
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA