Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
Artif Organs ; 33(9): 714-26, 2009 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-19775263

RESUMO

Design of a blood pump with a magnetically levitated rotor requires rigorous evaluation of the magnetic bearing and motor requirements and analysis of rotor dynamics and hydraulic performance with attention to hemolysis and thrombosis potential. Given the desired geometric dimensions, the required operating speed, flow in both the main and wash flow regions, and magnetic bearing performance, one of several design approaches was selected for a new prototype. Based on the estimated operating speed and clearance between the rotor and stator, the motor characteristics and dimensions were estimated. The motor stiffness values were calculated and used along with the hydraulic loading due to the fluid motion to determine the best design for the axial and radial magnetic bearings. Radial and axial stability of the left ventricular assist device prototype was verified using finite element rotor dynamic analysis. The analysis indicated that the rotor could be completely levitated and spun to the desired operating speed with low power loss and no mechanical contact. In vitro experiments with a mock loop test setup were performed to evaluate the performance of the new blood pump prototype.


Assuntos
Centrifugação , Coração Auxiliar , Magnetismo , Desenho de Equipamento , Análise de Elementos Finitos , Coração Auxiliar/efeitos adversos , Hemólise , Hemorreologia , Teste de Materiais , Modelos Cardiovasculares , Trombose/sangue , Trombose/etiologia , Trombose/prevenção & controle
2.
Artif Organs ; 32(5): 366-75, 2008 May.
Artigo em Inglês | MEDLINE | ID: mdl-18471166

RESUMO

The MiTiHeart (MiTiHeart Corporation, Gaithersburg, MD, USA) left ventricular assist device (LVAD), a third-generation blood pump, is being developed for destination therapy for adult heart failure patients of small to medium frame that are not being served by present pulsatile devices. The pump design is based on a novel, patented, hybrid passive/active magnetic bearing system with backup hydrodynamic thrust bearing and exhibits low power loss, low vibration, and low hemolysis. Performance of the titanium alloy prototype was evaluated in a series of in vitro tests with blood analogue to map out the performance envelop of the pump. The LVAD prototype was implanted in a calf animal model, and the in vivo pump performance was evaluated. The animal's native heart imparted a strong pulsatility to the flow rate. These tests confirmed the efficacy of the MiTiHeart LVAD design and confirmed that the pulsatility does not adversely affect the pump performance.


Assuntos
Fenômenos Eletromagnéticos/instrumentação , Coração Auxiliar , Teste de Materiais/instrumentação , Desenho de Prótese/instrumentação , Ligas/química , Animais , Materiais Biocompatíveis/química , Materiais Biomiméticos/química , Pressão Sanguínea/fisiologia , Temperatura Corporal/fisiologia , Bovinos , Falha de Equipamento , Glicerol/química , Hemólise/fisiologia , Masculino , Modelos Animais , Fluxo Pulsátil/fisiologia , Rotação , Titânio/química , Vibração , Viscosidade , Água/química
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA