Long-term clinical experience with Amplatzer Ductal Occluder II for closure of the persistent arterial duct in children.
Kang, Sok-Leng; Morgan, Gareth; Forsey, Jonathan; Tometzki, Andrew; Martin, Robin.
Afiliação
  • Kang SL; Bristol Congenital Heart Centre, Bristol Royal Hospital for Children and Bristol Royal Infirmary, University Hospitals Bristol, NHS Foundation Trust, Bristol, United Kingdom.
Catheter Cardiovasc Interv ; 83(7): 1102-8, 2014 Jun 01.
Article em En | MEDLINE | ID: mdl-24403100
ABSTRACT

OBJECTIVES:

To describe the long-term clinical experience and follow-up with the Amplatzer Ductal Occluder II (ADO II) in children.

METHODS:

All patients undergoing attempted transcatheter closure of patent arterial duct (PDA) with the ADO II were included. Data collected included demographic, clinical, and echocardiographic parameters.

RESULTS:

From March 2008 until March 2013, 62 patients with a median age of 1.2 years (range 0.43-11.1 years) and median weight of 9 kg (range 4.7-31.4 kg) underwent the procedure. The median measurement for minimal ductal diameter was 2.7 mm (range 1.3-5 mm). An ADO II was implanted in 60 patients (96.8%). Two patients had significant residual shunting following deployment of the ADO II and underwent closure with the Amplatzer ductal occluder I (ADO I) during the same procedure. In six patients, the initial ADO II was unsatisfactory, and after recapture a different size ADO II was deployed. Device embolization of the ADO II to the pulmonary artery occurred in 6.7% of patients. Of these, one underwent surgical closure and three were closed with an ADO I. Complete occlusion on echocardiography was noted prior to discharge in 87.5% of the deployed occluders and 100% at first follow-up. Five year follow-up (n = 25) revealed a 100% occlusion rate. There were three cases of persistent mild left pulmonary artery stenosis at long-term follow-up.

CONCLUSIONS:

The ADO II is effective for occlusion of PDA with variable anatomy from either arterial or venous approaches with a low profile delivery system. Stable occluder position is highly dependent on accurate device sizing, good quality imaging to visualize device configuration after deployment and operator experience.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Permeabilidade do Canal Arterial / Dispositivo para Oclusão Septal / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Permeabilidade do Canal Arterial / Dispositivo para Oclusão Septal / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Reino Unido