In Vitro Characterization of the Pittsburgh Pediatric Ambulatory Lung.
Orizondo, Ryan A; May, Alexandra G; Madhani, Shalv P; Frankowski, Brian J; Burgreen, Greg W; Wearden, Peter D; Federspiel, William J.
Afiliação
  • Orizondo RA; From the McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • May AG; From the McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • Madhani SP; Department of Chemical and Petroleum Engineering, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • Frankowski BJ; From the McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • Burgreen GW; Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • Wearden PD; From the McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, Pennsylvania.
  • Federspiel WJ; Computational Fluid Dynamics Group, Center for Advanced Vehicular Systems, Mississippi State University, Starkville, Mississippi.
ASAIO J ; 64(6): 806-811, 2018.
Article em En | MEDLINE | ID: mdl-29240630
ABSTRACT
Acute and chronic respiratory failure are a significant source of pediatric morbidity and mortality. Current respiratory support options used to bridge children to lung recovery or transplantation typically render them bedridden and can worsen long-term patient outcomes. The Pittsburgh Pediatric Ambulatory Lung (P-PAL) is a wearable pediatric blood pump and oxygenator (0.3 m surface area) integrated into a single compact unit that enables patient ambulation. The P-PAL is intended for long-term use and designed to provide up to 90% of respiratory support in children weighing 5-25 kg. Computational fluid dynamics and numerical gas exchange modeling were used to design the P-PAL and predict its performance. A P-PAL prototype was then used to obtain pressure versus flow curves at various impeller rotation rates using a blood analog fluid. In vitro oxygen exchange rates were obtained in blood in accordance with ISO standard 7199. The normalized index of hemolysis (NIH) was measured over a 6 hour period at blood flow rates of 1 and 2.5 L/min. The P-PAL provided blood flows of 1-2.5 L/min against the pressure drop associated with its intended-use pediatric cannulas. The oxygen exchange rate reached a maximum of 108 ml/min at a blood flow rate of 2.5 L/min and met our respiratory support design target. Device-induced hemolysis was low with NIH values of 0.022-0.027 g/100 L in the intended blood flow rate range. In conclusion, the current P-PAL design met our pumping, oxygenation, and hemolysis specifications and has the potential to improve treatment for pediatric respiratory failure.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração Artificial / Insuficiência Respiratória / Desenho de Equipamento Tipo de estudo: Prognostic_studies Limite: Child / Humans Idioma: En Revista: ASAIO J Assunto da revista: TRANSPLANTE Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração Artificial / Insuficiência Respiratória / Desenho de Equipamento Tipo de estudo: Prognostic_studies Limite: Child / Humans Idioma: En Revista: ASAIO J Assunto da revista: TRANSPLANTE Ano de publicação: 2018 Tipo de documento: Article