Nanoparticle Enhancement Cascade for Sensitive Multiplex Measurements of Biomarkers in Complex Fluids with Surface Plasmon Resonance Imaging.
Hendriks, Jan; Stojanovic, Ivan; Schasfoort, Richard B M; Saris, Daniël B F; Karperien, Marcel.
Afiliação
  • Hendriks J; Department of Developmental BioEngineering, MIRA Institute for Biomedical Technology and Technical Medicine , University of Twente , Enschede , 7522 NB , The Netherlands.
  • Stojanovic I; Medical Cell Biophysics, MIRA Institute for Biomedical Technology and Technical Medicine , University of Twente , Enschede , 7522 NB , The Netherlands.
  • Schasfoort RBM; Medical Cell Biophysics, MIRA Institute for Biomedical Technology and Technical Medicine , University of Twente , Enschede , 7522 NB , The Netherlands.
  • Saris DBF; Department of Orthopedics , UMC Utrecht , Utrecht , 3584 CX , The Netherlands.
  • Karperien M; Department of Reconstructive Medicine, MIRA Institute for Biomedical Technology and Technical Medicine, Faculty of Science and Technology , University of Twente , Enschede , 7522 NB , The Netherlands.
Anal Chem ; 90(11): 6563-6571, 2018 06 05.
Article em En | MEDLINE | ID: mdl-29732889
ABSTRACT
There is a large unmet need for reliable biomarker measurement systems for clinical application. Such systems should meet challenging requirements for large scale use, including a large dynamic detection range, multiplexing capacity, and both high specificity and sensitivity. More importantly, these requirements need to apply to complex biological samples, which require extensive quality control. In this paper, we present the development of an enhancement detection cascade for surface plasmon resonance imaging (SPRi). The cascade applies an antibody sandwich assay, followed by neutravidin and a gold nanoparticle enhancement for quantitative biomarker measurements in small volumes of complex fluids. We present a feasibility study both in simple buffers and in spiked equine synovial fluid with four cytokines, IL-1ß, IL-6, IFN-γ, and TNF-α. Our enhancement cascade leads to an antibody dependent improvement in sensitivity up to 40 000 times, resulting in a limit of detection as low as 50 fg/mL and a dynamic detection range of more than 7 logs. Additionally, measurements at these low concentrations are highly reliable with intra- and interassay CVs between 2% and 20%. We subsequently showed this assay is suitable for multiplex measurements with good specificity and limited cross-reactivity. Moreover, we demonstrated robust detection of IL-6 and IL-1ß in spiked undiluted equine synovial fluid with small variation compared to buffer controls. In addition, the availability of real time measurements provides extensive quality control opportunities, essential for clinical applications. Therefore, we consider this method is suitable for broad application in SPRi for multiplex biomarker detection in both research and clinical settings.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Citocinas / Ressonância de Plasmônio de Superfície / Nanopartículas Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: Anal Chem Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Citocinas / Ressonância de Plasmônio de Superfície / Nanopartículas Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: Anal Chem Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda