Your browser doesn't support javascript.
loading
In-Clinic Evaluation of the MiniMed 670G System "Suspend Before Low" Feature in Children with Type 1 Diabetes.
Wood, Michael A; Shulman, Dorothy I; Forlenza, Gregory P; Bode, Bruce W; Pinhas-Hamiel, Orit; Buckingham, Bruce A; Kaiserman, Kevin B; Liljenquist, David R; Bailey, Timothy S; Shin, John; Huang, Suiying; Chen, Xiaoxiao; Cordero, Toni L; Lee, Scott W; Kaufman, Francine R.
Afiliação
  • Wood MA; 1 Division of Pediatric Endocrinology, University of Michigan Medical School , Ann Arbor, Michigan.
  • Shulman DI; 2 USF Diabetes Center, Morsani College of Medicine, University of South Florida , Tampa, Florida.
  • Forlenza GP; 3 Barbara Davis Center for Childhood Diabetes , Aurora, Colorado.
  • Bode BW; 4 Atlanta Diabetes Associates , Atlanta, Georgia .
  • Pinhas-Hamiel O; 5 Edmond and Lily Safra Children's Hospital, Sheba Medical Center , Tel Aviv, Israel .
  • Buckingham BA; 6 Department of Pediatric Endocrinology, Stanford University , Stanford, California.
  • Kaiserman KB; 7 SoCal Diabetes , Torrance, California.
  • Liljenquist DR; 8 Rocky Mountain Diabetes and Osteoporosis Center , Idaho Falls, Idaho.
  • Bailey TS; 9 AMCR Institute , Escondido, California.
  • Shin J; 10 Medtronic , Northridge, California.
  • Huang S; 10 Medtronic , Northridge, California.
  • Chen X; 10 Medtronic , Northridge, California.
  • Cordero TL; 10 Medtronic , Northridge, California.
  • Lee SW; 10 Medtronic , Northridge, California.
  • Kaufman FR; 10 Medtronic , Northridge, California.
Diabetes Technol Ther ; 20(11): 731-737, 2018 11.
Article em En | MEDLINE | ID: mdl-30299976
ABSTRACT

BACKGROUND:

The Medtronic predictive low-glucose management (PLGM) algorithm automatically stops insulin delivery when sensor glucose (SG) is predicted to reach or fall below a preset low-glucose value within the next 30 min, and resumes delivery after hypoglycemia recovery. The present study evaluated the PLGM algorithm performance of the MiniMed™ 670G system SmartGuard™ "suspend before low" feature in children aged 7-13 years with type 1 diabetes (T1D).

METHOD:

Participants (N = 105, mean ± standard deviation of 10.8 ± 1.8 years) underwent an overnight in-clinic evaluation of the "suspend before low" feature with a preset low limit of 65 mg/dL. After exercise, frequent sample testing (FST) was conducted every 5 min if values were <70 mg/dL; every 15 min if 70-80 mg/dL; and every 30 min if >80 mg/dL. First-day performance of the Guardian™ Sensor 3 glucose sensor and continuous glucose monitoring system was also evaluated.

RESULTS:

Activation of the "suspend before low" feature occurred in 79 of the 105 participants, 79.7% (63/79) did not result in SG falling below 65 mg/dL. Mean glucose at activation was 102 ± 19 mg/dL and the initial insulin suspension duration was 87.5 ± 32.7 min. Four hours after insulin resumption, mean reference glucose was 130 ± 42 mg/dL. Mean absolute relative difference between the FST reference glucose and SG values on the first day of sensor wear was 11.4%. For the 26 participants in whom the "suspend before low" feature did not activate, none involved a reference glucose value ≤65 mg/dL, suggesting that the PLGM algorithm performed as intended.

CONCLUSION:

In children aged 7-13 years with T1D, the "suspend before low" feature of the MiniMed 670G system demonstrated a hypoglycemia prevention rate of nearly 80% after exercise and did not involve rebound hyperglycemia. There were no events of severe hypoglycemia during the evaluation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Algoritmos / Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Hipoglicemia / Hipoglicemiantes / Insulina Tipo de estudo: Etiology_studies / Evaluation_studies / Prognostic_studies Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Diabetes Technol Ther Assunto da revista: ENDOCRINOLOGIA / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Algoritmos / Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Hipoglicemia / Hipoglicemiantes / Insulina Tipo de estudo: Etiology_studies / Evaluation_studies / Prognostic_studies Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Diabetes Technol Ther Assunto da revista: ENDOCRINOLOGIA / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article