Marker-less assessment of the geometric error of fused cone-beam computed tomography images of the foot constructed using stitching software.
Wellenberg, Ruud Hh; Dobbe, Johannes Gg; Erkkilä, Jukka; Maas, Mario; Streekstra, Geert J.
Afiliação
  • Wellenberg RH; Department of Radiology and Nuclear Medicine, Amsterdam UMC, Location AMC, Amsterdam, The Netherlands.
  • Dobbe JG; Department of Biomedical Engineering and Physics, Amsterdam UMC, Location AMC, Amsterdam, The Netherlands.
  • Erkkilä J; Planmed Oy, Helsinki, Finland.
  • Maas M; Department of Radiology and Nuclear Medicine, Amsterdam UMC, Location AMC, Amsterdam, The Netherlands.
  • Streekstra GJ; Department of Radiology and Nuclear Medicine, Amsterdam UMC, Location AMC, Amsterdam, The Netherlands.
Acta Radiol ; 62(10): 1341-1348, 2021 Oct.
Article em En | MEDLINE | ID: mdl-33040567
BACKGROUND: Image noise, object repositioning, initial manual image alignment, and stitching of different volumes, i.e. anatomical regions may all affect the stitching error of fused cone-beam computed tomography (CBCT) images. PURPOSE: To determine the geometric error of fused CBCT images of the hindfoot, lower leg, and forefoot after using stitching software, based on a marker-less validation method. MATERIAL AND METHODS: CBCT images of the hindfoot, lower leg, and forefoot were acquired multiple times on a Planmed Verity scanner with and without repositioning the leg between acquisitions. Prototype stitching software was used to stitch hindfoot-forefoot volumes and hindfoot-lower leg volumes. Stitching error was determined via registration and by calculating the displacement of the tibia, first metatarsal, or proximal phalanges, with respect to the calcaneus, compared to their position on conventional CT. RESULTS: Overall total translation and rotation errors were 1.22 ± 0.62 mm (range 0.43-3.07 mm) and 1.27° ± 0.53° (range 0.29°-2.88°). Lower leg translation was smaller compared to forefoot translation (P < 0.001). No statistical difference was observed between lower leg and forefoot rotation errors (P = 0.186). Cadaver repositioning between acquisitions resulted in larger rotation errors (P < 0.05). Cadaver repositioning did not affect translation errors (P = 0.768). CONCLUSION: Geometric error of fused CBCT images can be quantified using a marker-less validation method. Stitching of hindfoot volumes with forefoot and lower leg volumes induces translation and rotation errors of up to approximately 3 mm and 3°. Translation errors are larger in hindfoot-forefoot stitches compared to hindfoot-lower leg stitches.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada de Feixe Cônico / Pé Limite: Humans Idioma: En Revista: Acta Radiol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada de Feixe Cônico / Pé Limite: Humans Idioma: En Revista: Acta Radiol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Holanda