Your browser doesn't support javascript.
loading
Corneal stromal roughness after VisuMax and Intralase femtosecond laser photodisruption: An atomic force microscopy study.
Gros-Otero, Juan; Ketabi, Samira; Cañones-Zafra, Rafael; Garcia-Gonzalez, Montserrat; Villa-Collar, Cesar; Casado, Santiago; Teus, Miguel A.
Afiliação
  • Gros-Otero J; Clínica Rementería, Madrid, Spain.
  • Ketabi S; Universidad CEU San Pablo, Madrid, Spain.
  • Cañones-Zafra R; Hospital Universitario Príncipe de Asturias, Alcalá de Henares, Madrid, Spain.
  • Garcia-Gonzalez M; Hospital Universitario Príncipe de Asturias, Alcalá de Henares, Madrid, Spain.
  • Villa-Collar C; Clínica Rementería, Madrid, Spain.
  • Casado S; Clínica Novovisión, Madrid, Spain.
  • Teus MA; Clínica Novovisión, Madrid, Spain.
PLoS One ; 16(5): e0252449, 2021.
Article em En | MEDLINE | ID: mdl-34043738
ABSTRACT

PURPOSE:

To compare the induced corneal stromal bed roughness measured with atomic force microscopy (AFM) after LASIK flap creation with the IntraLase 60 kHz and the VisuMax femtosecond laser platforms.

METHODS:

Three freshly enucleated porcine eyes were operated with each femtosecond laser in this experimental study. Standard LASIK treatment parameters were used for the experiment. After LASIK flap creation, the corneal stromal roughness was assessed using a JPK NanoWizard II® AFM in contact mode immersed in liquid. Olympus OMCL-RC800PSA commercial silicon nitride cantilever tips were used. Surface measurements were made in 10 regions of the central cornea of each sample measuring 20 x 20 microns, at 512 x 512 point resolution. Roughness was measured using the root-mean-square (RMS) value within the given regions.

RESULTS:

Measurements from 30 regions of the 3 eyes (10 measurements per eye) in the Intralase (FS1) group, and 30 regions of the 3 eyes (10 measurements per eye) in the VisuMax (FS2) group were analyzed. There was a statistically significant difference in mean ± standard deviation RMS values between the FS1 and the FS2 groups (360 ± 120 versus 230 ± 100 nm respectively; P< 0.00001).

CONCLUSION:

This AFM study indicates that the surface of the stromal bed after LASIK flap creation is smoother in the FS2 group than the FS1 group.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Microscopia de Força Atômica / Substância Própria / Ceratomileuse Assistida por Excimer Laser In Situ Limite: Animals Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Microscopia de Força Atômica / Substância Própria / Ceratomileuse Assistida por Excimer Laser In Situ Limite: Animals Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Espanha