Your browser doesn't support javascript.
loading
Lead extraction of an unintentional pacemaker lead implantation in the left ventricle from the left subclavian artery.
Strubé, Camille; Domain, Guillaume; Dognin, Nicolas; Dagenais, François; Plourde, Benoit; Jacques, Frédéric; Gilbert, Philippe; Steinberg, Christian; Philippon, François.
Afiliação
  • Strubé C; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
  • Domain G; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
  • Dognin N; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
  • Dagenais F; Institut universitaire de cardiologie et de pneumologie de Québec, Cardiac Surgery Division, Québec, Canada.
  • Plourde B; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
  • Jacques F; Institut universitaire de cardiologie et de pneumologie de Québec, Cardiac Surgery Division, Québec, Canada.
  • Gilbert P; Centre Hospitalier Universitaire de Québec, Laval University, Quebec, Canada.
  • Steinberg C; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
  • Philippon F; Institut universitaire de cardiologie et de pneumologie de Québec, Electrophysiology Division, Québec, Canada.
Pacing Clin Electrophysiol ; 45(2): 281-284, 2022 Feb.
Article em En | MEDLINE | ID: mdl-34730845
INTRODUCTION: Inadvertent implantation from the subclavian/axillary arteries is uncommon and has been rarely reported. Complications are frequent, mainly driven by thromboembolic events. Lead extraction, depending on the dwell time may be challenging and associated with thromboembolic events Case Report: We report the case of an 87-year-old man who had involuntarily pacemaker lead implanted into the left ventricle through the left subclavian artery. An extraction procedure was performed with surgical dissection to the artery and simple traction of the probe without complication. CONCLUSION: Percutaneous lead extraction using surgical dissection is a therapeutic option for less than one year old left ventricle inadvertent lead implantation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Artéria Subclávia / Erros Médicos / Remoção de Dispositivo / Ventrículos do Coração Limite: Aged80 / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Artéria Subclávia / Erros Médicos / Remoção de Dispositivo / Ventrículos do Coração Limite: Aged80 / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Canadá