Sensor-guided transcatheter aortic valve replacement.
J Invasive Cardiol
; 36(2)2024 Feb.
Article
em En
| MEDLINE
| ID: mdl-38335504
ABSTRACT
OBJECTIVES:
The SavvyWire(OpSens Medical) is a support wire for transcatheter aortic valve replacement (TAVR) procedures that, in addition to its dedicated left ventricle (LV) pacing capabilities, has a distal pressure sensor that measures live transvalvular hemodynamics during the procedure. We aimed to determine the safety, efficacy, and functionality of the SavvyWire during TAVR procedures in an all-comer population.METHODS:
We performed a multicentric, prospective, observational, single-arm, all-comers registry of patients with symptomatic, severe aortic stenosis undergoing TAVR in 3 Canadian centers. Data were collected in a dedicated database, and pre-specified questionnaires were fulfilled by the heart team implanters after each procedure.RESULTS:
A total of 60 patients were included (mean age 78.6 ± 7.2 years; 51% women; mean Society of Thoracic Surgeons score 2.2 ± 1.6%). TAVR was performed through a transfemoral approach in 90% of cases, and balloon- and self-expandable valves were used in 73% and 27% of patients, respectively. There were no cases of LV perforation, guidewire deformation, significant loss of capture, or major software malfunction. The rate of successful delivery of the TAVR system was 100%, and effective LV pacing was achieved in 98% of patients. The pre-TAVR mean gradient was 39 ± 14 mm Hg while the final post-TAVR gradient was 8 ± 5 mm Hg; the mean aortic systolic pressure during rapid pacing was 54 ± 12 mm Hg. In 97% of the TAVR procedures, SavvyWire's functionality was reported to be better or similar to other TAVR workhorse support wires.CONCLUSIONS:
SavvyWire was safe, effective, and functional for live transvalvular hemodynamic evaluation and rapid pacing during TAVR procedures. More studies with larger sample sizes and comparison against different wires and gradient measurement methods are warranted.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Estenose da Valva Aórtica
/
Próteses Valvulares Cardíacas
/
Substituição da Valva Aórtica Transcateter
Tipo de estudo:
Clinical_trials
/
Diagnostic_studies
/
Etiology_studies
/
Observational_studies
/
Qualitative_research
/
Risk_factors_studies
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
País/Região como assunto:
America do norte
Idioma:
En
Revista:
J Invasive Cardiol
Assunto da revista:
CARDIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Canadá