Development of a time-resolved fluorescence immunoassay kit for detecting canine coronavirus and parvovirus through double labeling.
Li, Laiqing; Chen, Cuicui; Liang, Huankun; Dong, Wenqi; Leontiev, V N; Voytov, Igor Vitalievich.
Afiliação
  • Li L; Belarusian State Technological University, 13a Sverdlov Str, 220006, Minsk, Minsk, Belarus.
  • Chen C; Guangzhou Youdi Bio-technology Co., Ltd, 510663, Guangzhou, China.
  • Liang H; Guangzhou Youdi Bio-technology Co., Ltd, 510663, Guangzhou, China.
  • Dong W; Guangzhou Zhenda Biopharmaceutical Technology Co., Ltd, 510663, Guangzhou, China.
  • Leontiev VN; Belarusian State Technological University, 13a Sverdlov Str, 220006, Minsk, Minsk, Belarus.
  • Voytov IV; Belarusian State Technological University, 13a Sverdlov Str, 220006, Minsk, Minsk, Belarus. rector@belstu.by.
Virol J ; 21(1): 64, 2024 03 11.
Article em En | MEDLINE | ID: mdl-38468354
ABSTRACT

OBJECTIVE:

Canine enteric coronavirus (CCV) and canine parvovirus type 2 (CPV-2) are the main pathogens responsible for acute gastroenteritis in dogs, and both single and mixed infections are common. This study aimed to establish a double-labeling time-resolved fluorescence immunoassay (TRFIA) to test and distinguish CCV and CPV-2 diseases.

METHODS:

A sandwich double-labeling TRFIA method was established and optimized using europium(III) (Eu3+)/samarium(III) (Sm3+) chelates. CCV/CPV-2 antigens were first captured by the immobilized antibodies. Then, combined with Eu3+/Sm3+-labeled paired antibodies, the Eu3+/Sm3+ fluorescence values were detected after dissociation to calculate the CCV/CPV-2 ratios. The performance, clinical performance and methodology used for laboratory (sensitivity, specificity, accuracy and stability) testing were evaluated.

RESULTS:

A double-label TRFIA for CCV and CPV-2 detection was optimized and established. The sensitivity of this TRFIA kit was 0.51 ng/mL for CCV and 0.80 ng/mL for CPV-2, with high specificity for CCV and CPV-2. All the accuracy data were less than 10%, and the recovery ranged from 101.21 to 110.28%. The kits can be temporarily stored for 20 days at 4 °C and can be stored for 12 months at temperatures less than - 20 °C. Based on a methodology comparison of 137 clinically suspected patients, there was no statistically significant difference between the TRFIA kit and the PCR method. Additionally, for CCV detection, the clinical sensitivity was 95.74%, and the clinical specificity was 93.33%. For CPV-2 detection, the clinical sensitivity was 92.86%, and the clinical specificity was 96.97%.

CONCLUSION:

In this study, a double-label TRFIA kit was prepared for CCV and CPV-2 detection with high laboratory sensitivity, specificity, accuracy, stability, clinical sensitivity and specificity. This kit provides a new option for screening/distinguishing between CCV and CPV-2 and may help improve strategies to prevent and control animal infectious diseases in the future.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parvovirus Canino / Coronavirus Canino / Infecções por Parvoviridae / Doenças do Cão Limite: Animals / Humans Idioma: En Revista: Virol J Assunto da revista: VIROLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Belarus

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parvovirus Canino / Coronavirus Canino / Infecções por Parvoviridae / Doenças do Cão Limite: Animals / Humans Idioma: En Revista: Virol J Assunto da revista: VIROLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Belarus