1.
Radiat Prot Dosimetry ; 198(7): 386-392, 2022 Jun 09.
Artículo en Inglés | MEDLINE | ID: mdl-35512690

RESUMEN

The aim of this study is to evaluate the personal dose equivalent Hp(10) in the most frequent (non-cardiac) paediatric interventional radiology (PIR) procedures: central venous catheters (CVC), hepatic/biliary and sclerotherapy interventions. i2 active solid-state dosemeters placed over the lead apron were used to monitor the exposure of three interventional radiologists over 18 months. A database was created to register all procedures performed by each radiologist (including the type of procedure and the kerma-area product, PKA). The mean Hp(10) per procedure for CVC, sclerotherapy and hepatic/biliary interventions was respectively 0.01 ± 0.01 mSv, 0.18 ± 0.13 mSv and 0.12 ± 0.06 mSv (k = 2). A similar value of Hp(10)/PKA was found despite the type of procedure or the patient weight (~10 µSv/Gy·cm2). There was high variability among individual interventions, probably due to the variable level of complexity, which led to uncertainties in the measurements' mean higher than those associated with the dosemeter's angular and energy dependence. i2 therefore proved suitable for monitoring Hp(10) in PIR procedures.


Asunto(s)
Exposición Profesional , Protección Radiológica , Niño , Humanos , Exposición Profesional/análisis , Ropa de Protección , Dosis de Radiación , Protección Radiológica/métodos , Radiología Intervencionista/métodos
2.
Med Phys ; 48(4): 1956-1966, 2021 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-33544901

RESUMEN

PURPOSE: To assess occupational lens exposure in a mixed interventional radiology department, comparing pediatric and adult procedures. To analyze the correlation between the lens dose and the doses measured at the chest and collar level and the kerma-area product (PKA ). METHODS: For 17 months, three radiologists performing both pediatric and adult interventions were monitored by means of 14 dosimeters per worker: 12 single-point optically stimulated luminescent (OSL) dosimeters calibrated in terms of Hp (0.07) were placed on the inside and outside of two pairs of lead glasses, one for pediatric procedures and one for adult interventions; another whole-body OSL dosimeter calibrated in terms of Hp (10) was placed over the thyroid shield; finally, an additional active solid-state dosimeter, also calibrated for Hp (10), was worn on the chest, over the apron. Furthermore, a database was created to register the demographic and dosimetric data of the procedures, as well as the name of the radiologist acting as first operator. RESULTS: For the three radiologists, who performed 276-338 procedures/year (20% pediatric), cumulative annual doses to the left bare eye exceeded 20 mSv (21-61 mSv). Considering the glasses' protection, annual doses exceeded 6 mSv (13-48 mSv) for both eyes. No important differences were observed in lens dose per procedure between pediatric and adult interventions (0.16 vs 0.18, 0.12 vs 0.09, and 0.07 vs 0.07 mSv), although lens dose per PKA was 4.1-4.5 times higher in pediatrics (5.8 vs 1.3, 3.3 vs 0.8, and 2.6 vs 0.6 µSv/Gy·cm2 ) despite a similar use of the ceiling-suspended screen. Lens doses were highly correlated with collar readings (with Pearson coefficients [r] ranging from 0.86 to 0.98) and with chest readings (with r ranging from 0.75 to 0.93). However, slopes of the linear regressions varied greatly among radiologists. CONCLUSIONS: There is real risk of exceeding the occupational dose limit to the eye lens in mixed interventional radiology rooms if radiation protection tools are not used properly. Regular monitoring of the lens dose is recommended, given lens exposure might easily exceed 6 mSv/yr. Using a collar dosimeter for this purpose might be suitable if it is preceded by an individualized regression analysis. The same radiation protection measures should be applied to interventional radiologists regardless of whether they are treating pediatric or adult patients.


Asunto(s)
Cristalino , Exposición Profesional , Pediatría , Exposición a la Radiación , Adulto , Niño , Humanos , Exposición Profesional/análisis , Dosis de Radiación , Exposición a la Radiación/análisis , Radiología Intervencionista
3.
Phys Med ; 72: 1-6, 2020 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-32179406

RESUMEN

PURPOSE: To establish local diagnostic reference levels (DRLs) for non-cardiac interventional procedures in paediatrics. METHODS: The type of procedure, the patient's weight and age and dose-related data from 279 interventions was recorded in a database completed by interventional radiologists, radiographers and technicians of the Medical Physics department. These procedures were classified into 14 categories and 6 weight ranges. Local DRLs were proposed for those ranges in which a sample of at least 15 patients could be gathered and were calculated as the third quartile (Q3) of the air kerma-area product (PKA) values. The Q3 of the fluoroscopy time (FT) and number of digital subtraction angiography (DSA) images were also obtained. Finally, the correlation between PKA and weight was analysed. RESULTS: Local DRLs are proposed for three types of procedures: hepatic/biliary interventions (5-15 kg, 1304 cGy·cm2; 15-30 kg, 2121 cGy·cm2), sclerotherapy procedures (15-30 kg, 704 cGy·cm2; 30-50 kg, 4049 cGy·cm2; 50-80 kg, 3734 cGy·cm2) and central venous catheter (CVC) procedures (5-15 kg, 84 cGy·cm2). Hepatic/biliary interventions showed a moderate correlation (r = 0.61), while sclerotherapy procedures presented a poor correlation (r = 0.34) between PKA and weight, possibly due to the PKA dependence on the complexity level. Regarding CVC procedures, a clearly higher correlation was found when the fluoroscopy PKA value was normalised to the FT (r = 0.85 vs r = 0.35). CONCLUSIONS: The results support the feasibility of establishing DRLs for the most common procedures (sclerotherapy, hepatic/biliary and CVC interventions) despite the small number of paediatric interventions.


Asunto(s)
Radiología Intervencionista/normas , Peso Corporal , Humanos , Pediatría/normas , Estándares de Referencia
4.
J Appl Clin Med Phys ; 19(6): 124-132, 2018 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-30255659

RESUMEN

This work aims to study the effect on surface dose and dose distribution caused by the Elekta Fraxion cranial immobilization system. The effect of Fraxion inclusion in Elekta Monaco treatment planning system and its calculation accuracy is also checked. To study the dose attenuation, a cylindrical phantom was located over the Elekta Fraxion with an IBA CC13 ionization chamber placed in the central insert at the linac isocenter. Dose measurements at multiple gantry angles were performed for three open fields, 10 × 10 cm, 5 × 5 cm and other smaller 2 × 2 cm. Measured doses were compared with the ones calculated by Monaco. Surface dose and dose distribution in the buildup region were measured placing several Gafchromic Films EBT3 at linac CAX between the slabs of a RW3 phantom located over Fraxion and read using FilmQA Pro software. Measures were performed for two open field sizes and results were compared with Monaco calculations. Measurements show a 1% attenuation for 180° gantry angle but it can be as high as 3.4% (5 × 5 open field) for 150°/210° gantry angle, as with these angles the beam goes through the Fraxion's headrest twice. If Fraxion is not included in the calculation Monaco calculation can result in a 3% difference between measured and calculated doses, while with Fraxion in the calculation, the maximum difference is 0.9%. Fraxion increases 3.7 times the surface dose, which can be calculated by Monaco with a difference lower than 2%. Monaco also calculated correctly the PDD for both open fields (2%) when Fraxion is included in the calculation. This work shows that the attenuation varies with gantry angle. The inclusion of Fraxion in Monaco improves the calculation from 3% difference to 1% in the worst case. Furthermore, the surface dose increment and the dose in the buildup region are correctly calculated.


Asunto(s)
Cabeza , Método de Montecarlo , Neoplasias/radioterapia , Aceleradores de Partículas/instrumentación , Fantasmas de Imagen , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia de Intensidad Modulada/métodos , Algoritmos , Humanos , Radiometría/métodos , Dosificación Radioterapéutica
5.
Phys Med Biol ; 62(2): 633-651, 2017 01 21.
Artículo en Inglés | MEDLINE | ID: mdl-28044999

RESUMEN

Positron emission tomography (PET) images are characterised by low signal-to-noise ratio and blurred edges when compared with other image modalities. It is therefore advisable to use noise reduction methods for qualitative and quantitative analyses. Given the importance of the maximum and mean uptake values, it is necessary to avoid signal loss, which could modify the clinical significance. This paper proposes a method of non-linear image denoising for PET. It is based on spatially adaptive wavelet-shrinkage and uses context modelling, which explicitly considers the correlation between neighbouring pixels. This context modelling is able to maintain the uptake values and preserve the edges in significant regions. The algorithm is proposed as an alternative to the usual filtering that is performed after reconstruction.


Asunto(s)
Algoritmos , Aumento de la Imagen/métodos , Interpretación de Imagen Asistida por Computador/métodos , Fantasmas de Imagen , Tomografía de Emisión de Positrones/métodos , Relación Señal-Ruido , Humanos
6.
J Appl Clin Med Phys ; 17(4): 357-376, 2016 07 08.
Artículo en Inglés | MEDLINE | ID: mdl-27455474

RESUMEN

The aim of this study was to assess image quality and radiation dose of a biplane angiographic system with cone-beam CT (CBCT) capability tuned for pediatric cardiac procedures. The results of this study can be used to explore dose reduction techniques. For pulsed fluoroscopy and cine modes, polymethyl methacrylate phantoms of various thicknesses and a Leeds TOR 18-FG test object were employed. Various fields of view (FOV) were selected. For CBCT, the study employed head and body dose phantoms, Catphan 504, and an anthropomorphic cardiology phantom. The study also compared two 3D rotational angiography protocols. The entrance surface air kerma per frame increases by a factor of 3-12 when comparing cine and fluoroscopy frames. The biggest difference in the signal-to- noise ratio between fluoroscopy and cine modes occurs at FOV 32 cm because fluoroscopy is acquired at a 1440 × 1440 pixel matrix size and in unbinned mode, whereas cine is acquired at 720 × 720 pixels and in binned mode. The high-contrast spatial resolution of cine is better than that of fluoroscopy, except for FOV 32 cm, because fluoroscopy mode with 32 cm FOV is unbinned. Acquiring CBCT series with a 16 cm head phantom using the standard dose protocol results in a threefold dose increase compared with the low-dose protocol. Although the amount of noise present in the images acquired with the low-dose protocol is much higher than that obtained with the standard mode, the images present better spatial resolution. A 1 mm diameter rod with 250 Hounsfield units can be distinguished in reconstructed images with an 8 mm slice width. Pediatric-specific protocols provide lower doses while maintaining sufficient image quality. The system offers a novel 3D imaging mode. The acquisition of CBCT images results in increased doses administered to the patients, but also provides further diagnostic information contained in the volumetric images. The assessed CBCT protocols provide images that are noisy, but with very good spatial resolution.


Asunto(s)
Tomografía Computarizada de Haz Cónico/métodos , Cabeza/diagnóstico por imagen , Corazón/diagnóstico por imagen , Procesamiento de Imagen Asistido por Computador/normas , Fantasmas de Imagen , Angiografía/métodos , Fluoroscopía/métodos , Humanos , Procesamiento de Imagen Asistido por Computador/métodos , Imagenología Tridimensional , Polimetil Metacrilato/química , Dosis de Radiación , Relación Señal-Ruido