Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros












Base de datos
Intervalo de año de publicación
1.
Rev Sci Instrum ; 92(4): 043544, 2021 Apr 01.
Artículo en Inglés | MEDLINE | ID: mdl-34243392

RESUMEN

The x-ray imaging crystal spectrometer (XICS) is proposed as the principal method of diagnostics for plasma ion temperature and rotation for the China Fusion Engineering Test Reactor (CFETR) for its simplicity in implementation and no reliance on neutral beams. For D-T experiments with the electron temperature as high as 35-40 keV at the core region, highly charged high-Z ions can serve as the diagnostic ions for the XICS. For the CFETR, Xe44+, Xe51+, and W64+ are selected as the impurity ions. Appropriate crystal parameters are selected, as well as the preliminary layout for the spectrometer. We estimated the general performance of the spectrometer, including the emissivity of the impurities, the spatial resolution of the x-ray detector, and the expected count rate of line emissions. For the application in the fusion reactor environment, the effect of neutron irradiation on the crystal is briefly discussed.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...