Guang Pu Xue Yu Guang Pu Fen Xi ; 33(2): 571-4, 2013 Feb.
Artículo en Chino | MEDLINE | ID: mdl-23697157

RESUMEN

Retrieval of center wavelength and bandwidth is a key step for quantitative analysis of hyperspectral data. The present paper proposes a spectral calibration method of hyperspectral imager, whose spectrum covers visible and near-infrared band, using spectral absorption target. Ground calibration experiment was designed for a hyperspectral imager with a bandwidth of 6 nm. Hyperspectral imager and ASD spectrometer measured the same spectral absorption target synchronously. Reflectance spectrum was derived from the different data set. Center wavelength and bandwidth were retrieved by matching the reflectance data from hyperspectral imager and ASD spectrometer. The experiment result shows that this method can be applied in spectral calibration of hyperspectral imagers to improve the quantitative studies on hyperspectral imagery.

2.
Guang Pu Xue Yu Guang Pu Fen Xi ; 32(2): 430-4, 2012 Feb.
Artículo en Chino | MEDLINE | ID: mdl-22512184

RESUMEN

With the data in Urad Front Banner, Inner Mongolia on November 14th, 2010, hyper-spectral camera on UAV was calibrated adopting reflectance-based method. During the in-flight absolute radiometric calibration, 6 hyper-spectral radiometric gray-scale targets were arranged in the validation field. These targets' reflectances are 4.5%, 20%, 30%, 40%, 50% and 60% separately. To validate the calibration result, four extra hyper-spectral targets with sharp-edge spectrum were arranged to simulate the reflection and absorption peaks in natural objectives. With these peaks, the apparent radiance calculated by radiation transfer model and that calculated through calibration coefficients are much different. The result shows that in the first 15 bands (blue bands), errors are somewhat huge due to the noises of equipment. In the rest bands with quite even spectrum, the errors are small, most of which are less than 10%. For those bands with sharp changes in spectral curves, the errors are quite considerable, varying from 10% to 25%.

3.
Guang Pu Xue Yu Guang Pu Fen Xi ; 32(12): 3169-74, 2012 Dec.
Artículo en Chino | MEDLINE | ID: mdl-23427528

RESUMEN

Based on the data of the scientific experiment in Urad Front Banner for UAV Remote Sensing Load Calibration Field project, with the help of 6 hyperspectral radiometric targets with good Lambertian property, the wide-view multispectral camera in UAV was calibrated adopting reflectance-based method. The result reveals that for green, red and infrared channel, whose images were successfully captured, the linear correlation coefficients between the DN and radiance are all larger than 99%. In final analysis, the comprehensive error is no more than 6%. The calibration results demonstrate that the hyperspectral targets equipped by the calibration field are well suitable for air-borne multispectral load in-flight calibration. The calibration result is reliable and could be used in the retrieval of geophysical parameters.