Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros












Base de datos
Intervalo de año de publicación
1.
Rapid Commun Mass Spectrom ; 27(24): 2818-22, 2013 Dec 30.
Artículo en Inglés | MEDLINE | ID: mdl-24214868

RESUMEN

RATIONALE: An experimental Isobar Separator for Accelerator Mass Spectrometry (ISAMS) instrument has been used to demonstrate an on-line separation of HfF5(-) from its isobar WF5(-). This is necessary, in addition to sample preparation chemistry, for measuring (182)Hf at natural levels by Accelerator Mass Spectrometry (AMS). METHODS: The device utilizes a radiofrequency quadrupole (RFQ) controlled gas cell, wherein anion-gas reactions at eV energies attenuate the interfering isobars of the analyte molecular anions, leaving HfF5(-) for AMS analysis. The RFQ also helps to control the multiple scattering resulting from the ion-gas collisions. RESULTS: O2 gas was used in the HfF5(-)/WF5(-) separation and WF5(-) was attenuated by nearly 3 orders of magnitude while maintaining ~75% transmission of HfF5(-). It is expected that the transmission and attenuation can be increased by further research. CONCLUSIONS: This result advances the possibility of detecting natural (182)Hf when AMS is supplemented with an isobar separator in the injection system.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...