RESUMEN
Although total hip replacements (THR) can be considered one of the most successful implantable medical devices in history, wear remains the ultimate challenge in order to further increase clinical success. Wear assessment on retrieved implants is the most reliable way to perform research into failure mechanisms. Therefor the bearing surface of the explant is measured geometrically by coordinate measuring machine (CMM). Wear determination in geometrical data is carried out in 3 steps: (1) identifying the worn area, (2) reconstructing the pre-wear geometry and (3) quantify wear as the difference between worn area and pre-wear geometry. In previous studies, assumptions to pre-wear geometry had been made for wear determination (step 2) and the worn area was identified by deviations between measured data and assumed form. Thus, the original form of the retrieved endoprostheses, including form deviations due to the manufacturing process and implantation, was not considered which leads to uncertainties in the wear computed. This work introduces a method which allows to identify the wear area without making assumptions to the original form. Instead, the curvature of the bearing surface obtained by simple computations on the measurement data is analysed and the edge of the wear region is recognized by its deviation in curvature. The method is applied to a retrieved Metal-on-Metal prosthetic head and the results are compared to those of the well-known method introduced by Jaeger et al., in 2013. With the new approach the wear region is identified more accurately.