1.
Med Eng Phys ; 111: 103925, 2023 01.
Artículo en Inglés | MEDLINE | ID: mdl-36792249

RESUMEN

This study investigated the accuracy of the Insole3 wireless shoe device in estimating several clinically useful spatiotemporal parameters (STPs). Eleven subjects walked at slow (0.8-1.0 m/s) and moderate-paced (1.2-1.4 m/s) speeds. Data were simultaneously recorded using the Insole3 and an industry-standard, three-dimensional motion capture (MOCAP) system. An error analysis compared the resulting STP data from the two systems. The mean bias error (MBE) was generally lower for temporal variables, and somewhat higher, but acceptable, for spatial variables. The MBE for temporally-related cadence and cycle time were the lowest (less than ±0.45%), with 100% (110/110) of slow-paced walking trial values and 99.1% (109/110) of moderate-paced walking trial values within 5% of the MOCAP estimates. The MBE was highest for speed (3.23-4.91%) and stride length (3.68-4.63%), with between 52.7 and 69.1% of trial values falling within the 5% error range. Stance time and swing time ranged between -0.98 and 4.38% error for both walking conditions. The results of this study suggest that the Insole3 is a potential alternative to MOCAP for estimating several STPs, namely cadence, stance time, and cycle time, particularly for use outside of the laboratory setting.


Asunto(s)
Marcha , Caminata , Humanos , Fenómenos Biomecánicos
SELECCIÓN DE REFERENCIAS