Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros












Base de datos
Intervalo de año de publicación
1.
Clin Lab ; 69(8)2023 Aug 01.
Artículo en Inglés | MEDLINE | ID: mdl-37560846

RESUMEN

BACKGROUND: This study aimed to demonstrate potential measures to discern interference and establish an algorithm for identifying interference in an hs-cTnT assay with a clinical discrepancy. METHODS: Laboratory error and disease factors were first excluded. Then, a serial dilution test, polyethylene glycol (PEG) precipitation, and heterophile antibody blocking reagents (HBRs) were performed to confirm the interference substance. In addition, high-speed centrifugation and western blot were tried. RESULTS: The sample diluted linearly (R2 = 0.9966). However, there was a dramatic reduction in the concentration of hs-cTnT concentration to a normal level after both PEG precipitation and HBR1 treatment. The results were inconclusive in high-speed centrifugation and western blot assay. CONCLUSIONS: This study elucidated a heterophile antibody interference in measurement of hs-cTnT alone on a Roche analyzer.


Asunto(s)
Anticuerpos , Troponina T , Humanos , Indicadores y Reactivos , Western Blotting , Biomarcadores
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...