Med Sci Sports Exerc ; 56(6): 1196-1207, 2024 Jun 01.
Artículo en Inglés | MEDLINE | ID: mdl-38377012

RESUMEN

INTRODUCTION: Current wearables that collect heart rate and acceleration were not designed for children and/or do not allow access to raw signals, making them fundamentally unverifiable. This study describes the creation and calibration of an open-source multichannel platform (PATCH) designed to measure heart rate and acceleration in children ages 3-8 yr. METHODS: Children (N = 63; mean age, 6.3 yr) participated in a 45-min protocol ranging in intensities from sedentary to vigorous activity. Actiheart-5 was used as a comparison measure. We calculated mean bias, mean absolute error (MAE) mean absolute percent error (MA%E), Pearson correlations, and Lin's concordance correlation coefficient (CCC). RESULTS: Mean bias between PATCH and Actiheart heart rate was 2.26 bpm, MAE was 6.67 bpm, and M%E was 5.99%. The correlation between PATCH and Actiheart heart rate was 0.89, and CCC was 0.88. For acceleration, mean bias was 1.16 mg and MAE was 12.24 mg. The correlation between PATCH and Actiheart was 0.96, and CCC was 0.95. CONCLUSIONS: The PATCH demonstrated clinically acceptable accuracies to measure heart rate and acceleration compared with a research-grade device.


Asunto(s)
Frecuencia Cardíaca , Humanos , Niño , Preescolar , Frecuencia Cardíaca/fisiología , Masculino , Femenino , Calibración , Aceleración , Dispositivos Electrónicos Vestibles , Acelerometría/instrumentación
SELECCIÓN DE REFERENCIAS