J Pharm Biomed Anal ; 225: 115239, 2023 Feb 20.
Artículo en Inglés | MEDLINE | ID: mdl-36638567

RESUMEN

A direct reversed-phase high-performance liquid chromatographic (HPLC) method was developed for determining the content of the enantiomeric impurity of the chiral statin rosuvastatin calcium salt (RSV) in commercial tablets. The baseline enantioseparation was achieved using the Lux Cellulose-2 column and a binary linear gradient of acetonitrile and trifluoroacetic acid 0.05% in an aqueous solution. The flow rate of the mobile phases and column temperature were set at 1.0 mL min- 1 and 40 °C, respectively. In comparison with the isocratic HPLC method reported in the European Pharmacopoeia (EP) monograph for RSV, the gradient elution method offered improved chemo-and enantio-selectivity and reduced analysis times. The limits of quantitation and detection of the enantiomeric impurity were found to be 0.15 and 0.05 µg mL-1.


Asunto(s)
Celulosa , Agua , Rosuvastatina Cálcica , Celulosa/química , Cromatografía Líquida de Alta Presión/métodos , Estereoisomerismo , Agua/química