Catheter Cardiovasc Interv ; 62(2): 175-80, 2004 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-15170706

RESUMEN

We evaluated two different intravascular ultrasound (IVUS) systems, Atlantis and Intrafocus, to verify their accuracy and reproducibility. In an in vivo study on 20 consecutive patients with coronary artery diseases, the minimum lumen diameter (MLD), vessel diameter, lumen area (LA), vessel area, plaque area, and area stenosis rate (% AS) were respectively measured. In an in vitro study, MLD and LA were measured in four metal tubes with different diameters. All of the measured values except for % AS by Atlantis were significantly larger than the values obtained with Intrafocus. Nonuniform rotational distortion (NURD) was estimated as 34% in Atlantis and 1% in Intrafocus. The measurements by Atlantis were larger than the true values while the measurements by Intrafocus were less than the true values in all four metal tubes. These findings suggest that we should clearly avoid the use of different IVUS systems in the same study.


Asunto(s)
Enfermedad de la Arteria Coronaria/diagnóstico por imagen , Ultrasonografía Intervencional , Anciano , Calcinosis/diagnóstico por imagen , Calcinosis/epidemiología , Enfermedad de la Arteria Coronaria/epidemiología , Diseño de Equipo , Femenino , Humanos , Aumento de la Imagen , Incidencia , Masculino , Persona de Mediana Edad , Variaciones Dependientes del Observador , Sensibilidad y Especificidad , Estadística como Asunto