1.
Zhongguo Yi Liao Qi Xie Za Zhi ; 38(4): 259-60, 281, 2014 Jul.
Artículo en Chino | MEDLINE | ID: mdl-25330604

RESUMEN

This article describes the design and development of an ultrasonic power meter which is consist of an electronic balance, a practice target, an acoustic enclosures and a blocking. The electronic balance mounted on the blocking is linked with the practice target by connecting rod. By adjusting the blocking makes the practice target suspended above ultrasound probe, and then the ultrasonic power can be measured. After initial tests, the ultrasonic power meter performanced with good stability and high precision.


Asunto(s)
Diseño de Equipo , Ultrasonido
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...