Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros












Base de datos
Intervalo de año de publicación
1.
Appl Radiat Isot ; 134: 385-390, 2018 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-29248210

RESUMEN

An international key comparison, identifier CCRI(II)-K2.Ge-68, has been performed. The National Institute of Standards and Technology (NIST) served as the pilot laboratory, distributing aliquots of a 68Ge/68Ga solution. Results for the activity concentration, CA, of 68Ge at a reference date of 12h00 UTC 14 November 2014 were submitted by 17 laboratories, encompassing many variants of coincidence methods and liquid-scintillation counting methods. The first use of 4π(Cherenkov)ß-γ coincidence and anticoincidence methods in an international comparison is reported. One participant reported results by secondary methods only. Two results, both utilizing pure liquid-scintillation methods, were identified as outliers. Evaluation using the Power-Moderated Mean method results in a proposed Comparison Reference Value (CRV) of 621.7(11)kBqg-1, based on 14 results. The degrees of equivalence and their associated uncertainties are evaluated for each participant. Several participants submitted 3.6mL ampoules to the BIPM to link the comparison to the International Reference System (SIR) which may lead to the evaluation of a Key Comparison Reference Value and associated degrees of equivalence.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...