Pediatr Rep ; 16(1): 69-76, 2024 Jan 12.
Artículo en Inglés | MEDLINE | ID: mdl-38251316

RESUMEN

Umbilical venous (UV) catheters (UVCs) are commonly used in severely ill neonates. Complications associated with UVC often result from an inappropriate UVC position. Calcification of the UV, a rare complication, was observed in an extremely low-birth-weight infant born at 23 weeks of gestation. After birth, the infant experienced respiratory and circulatory dysfunction, followed by disseminated intravascular coagulation (DIC). A UVC was inserted, and circulatory agonists and blood transfusions were administered, as well as a calcium gluconate infusion for hypocalcemia and hyperkalemia. Ten days after birth, calcification was detected in the UV, likely due to a tunica intima injury caused by UVC, a hypercoagulable state due to DIC, and a high-dose calcium gluconate infusion. Additionally, proximal port malpositioning of the double-lumen catheter might have contributed to calcification within the UV. To prevent such complications, real-time ultrasound confirmation with agitated saline contrast during UVC placement is recommended; in the absence of the facility or skills for ultrasonography, X-rays should be performed in the lateral and anteroposterior views. Furthermore, when using multi-lumen catheters, physicians should not only verify the tip position but also ensure proper placement of proximal ports and carefully select medications administered through the ports.