BACKGROUND: In the past few decades, the inadequate reprocessing of bronchoscopes has been associated with several serious outbreaks caused by multidrug-resistant microorganisms. In this study we evaluated the improvement in the quality of reprocessing in a Bronchoscopy Unit (BU), after the introduction of a new procedure. METHODS: In 2019, observational and clinical audits were conducted in the BU. After the introduction of an improved procedure in 2020, a microbiological surveillance plan was implemented in 2021. RESULTS: In 2019, 13 of 22 bronchoscopes (59%) resulted as non-compliant, 18% as high concern organisms (HCO) and 36.4% as high microbial count (≥100 CFU/all channels) and HCO. The most frequent microorganisms were Staphylococcus aureus (38.5%) and NDM-producing Klebsiella pneumoniae (15.4%). The bronchoscopes were stored inside their transport cases, which in some cases were found to be contaminated by the same strains isolated on the bronchoscopes (Enterobacter gergoviae and Vibrio alginolyticus). In 2021, all 31 bronchoscopes were sampled at least three times and 13/99 (13.1%) resulted as non-compliant, mostly K. pneumoniae (4.04%). Contamination level increases weakly in bronchoscopes in use for more than 14 years (R = 0.32). CONCLUSIONS: The adoption of an improved reprocessing procedure decreased the non-compliance of bronchoscopes, increasing the quality of the process and patient safety.
Assuntos
Broncoscópios , Infecção Hospitalar , Broncoscópios/microbiologia , Infecção Hospitalar/epidemiologia , Desinfecção/métodos , Contaminação de Equipamentos/prevenção & controle , Klebsiella pneumoniaeRESUMO
Background: Correct reprocessing and microbiological surveillance on endoscopes are fundamental for preventing the transmission of multi-drug resistant strains and device-related infections. Methods: A questionnaire with three domains was created: (1) centre characteristics; (2) endoscope reprocessing procedures; and (3) application of microbiological surveillance. Nurses working in endoscopic units across Italy were invited to anonymously fill out the questionnaire on the SurveyMonkey platform between November 2021 and February 2022. Results: A total of 82 out of 132 endoscopic centres participated in the survey, with at least one centre from each Italian region. Data found different concerns regarding the current practice of both reprocessing and microbiological surveillance. According to respondents, the training on reprocessing was performed through theoretical training and only in 10% of centres; the microbiological surveillance was regularly performed in 59% of centres; and sampled endoscopes were not excluded for use in 31% of centres performing the surveillance until the outcome was pending, and when positive, 72% maintained them in quarantine until a successive negative result. Conclusions: Reprocessing and microbiological surveillance currently present several criticisms along the endoscopic centres in Italy. Our survey highlights the need for the correct application of the national recommendations in each endoscopic centre to prevent the potential transmission of endoscope-related infections.