A determination method of flomoxef (FMOX) concentration in serum by capillary electrophoresis is developed. Serum samples are extracted with acetonitrile. After pretreatment, they are separated in a fused-silica capillary tube with a 25 mM borate buffer (pH 10.0) as a running buffer that contains 50mM sodium dodecyl sulfate. The FMOX and acetaminophen (internal standard) are detected by UV absorbance at 200 nm. Linearity (0-200 mg/L) is good, and the minimum limit of detection is 1.0 mg/L (S/N = 3). The relative standard deviations of intra- and interassay variability are 1.60-4.78% and 2.10-3.31%, respectively, and the recovery rate is 84-98%. This method can be used for determination of FMOX concentration in serum.