Investigation on stick phenomena in metal-on-metal hip joints after resting periods.
Wimmer, M A; Nassutt, R; Sprecher, C; Loos, J; Täger, G; Fischer, A.
Afiliación
  • Wimmer MA; Department of Orthopedic Surgery, Rush University Medical Center, Chicago, Illinois 60612, USA. markus_a_wimmer@rush.edu
Proc Inst Mech Eng H ; 220(2): 219-27, 2006 Feb.
Article en En | MEDLINE | ID: mdl-16669389
ABSTRACT
Insufficient understanding of tribological behaviour in total joint arthroplasty is considered as one of the reasons for prosthesis failure. Contrary to the continuous motion input profiles of hip simulators, human locomotion contains motion interruptions. These occurring resting periods can cause stick phenomena in metal-on-metal hip joints. The aim of the present study was to investigate the tribological sensitivity of all-metal bearings to motion interruptions on in vitro test specimens and retrieved implants. Friction and wear with and without resting periods were quantified. Unlike the metal-on-polyethylene joints, the static friction of metal-on-metal joints increased up to micros = 0.3 with rest, while wear appeared to be unaffected. This effect is caused by the interlocking of firmly adhered carbon layers, which were generated from the protein-containing lubricant through tribochemical reactions. Since more than 80 per cent of the retrieved implants exhibited macroscopically visible carbon layers, the increase in friction presumably also occurs under physiological conditions, which is then transferred to the bone-implant interface. These recurrent tangential stress peaks should be considered for the design features of the cup-bone interface, in particular when larger-sized implant heads are used.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vitalio / Materiales Biocompatibles / Falla de Prótesis / Prótesis de Cadera Tipo de estudio: Evaluation_studies Idioma: En Revista: Proc Inst Mech Eng H Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vitalio / Materiales Biocompatibles / Falla de Prótesis / Prótesis de Cadera Tipo de estudio: Evaluation_studies Idioma: En Revista: Proc Inst Mech Eng H Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
...