Your browser doesn't support javascript.
loading
Günther Tulip filter retrievability multicenter study including CT follow-up: final report.
Hoppe, Hanno; Nutting, Charles W; Smouse, H Robert; Vesely, Thomas M; Pohl, Christoph; Bettmann, Michael A; Kaufman, John A.
Afiliación
  • Hoppe H; Dotter Interventional Institute, Oregon Health Science University, Portland 97239-3098, USA.
J Vasc Interv Radiol ; 17(6): 1017-23, 2006 Jun.
Article en En | MEDLINE | ID: mdl-16778236
ABSTRACT

PURPOSE:

To evaluate the safety and effectiveness of retrieval of the Günther Tulip inferior vena cava (IVC) filter. MATERIALS AND

METHODS:

This was a nonrandomized, single-armed, multicenter prospective investigation. Patients at temporary high risk for pulmonary embolism (PE) or deep vein thrombosis (DVT) who did not require a permanent filter were eligible. Forty-one patients received 42 Günther Tulip filters 22 men and 19 women with a mean age of 47.7 years. Indications for filter placement included prophylaxis, PE, and DVT. Three months after filter retrieval, contrast agent-enhanced computed tomography of the abdomen, jugular vein ultrasonography, and clinical follow-up were performed.

RESULTS:

The filter retrieval rate was 57% (23 of 41). Günther Tulip filters were removed at a mean of 11.1 days (range, 2-14 d). The technical and clinical success rates for filter retrieval were both 100%. One placement complication and two protocol deviations occurred. These patients were excluded in terms of retrieval-related outcomes. One case of PE occurred with a filter in place, and one filter migrated to the heart. There were no acute complications caused by filter retrieval. At 3-month follow-up, there was no recurrent PE, DVT, jugular vein occlusion, or IVC stenosis or occlusion.

CONCLUSION:

In this multicenter study, retrieval of the Günther Tulip filter was safe and without recurrent thromboembolic events or evidence of IVC or jugular vein damage at 3-month follow-up.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Embolia Pulmonar / Filtros de Vena Cava / Trombosis de la Vena / Remoción de Dispositivos Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Guideline / Observational_studies / Risk_factors_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Embolia Pulmonar / Filtros de Vena Cava / Trombosis de la Vena / Remoción de Dispositivos Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Guideline / Observational_studies / Risk_factors_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
...