Your browser doesn't support javascript.
loading
[New method for accurate ECG trigger of intraaortic balloon counterpulsation during off-pump coronary artery bypass surgery].
Mori, Yosuke; Kaneko, Takehiko; Aoyama, Yasuhiko; Suzuki, Seiko; Takayama, Wataru; Miyashita, Keiko; Yoshikawa, Tamotsu; Kobayasii, Yoshiro; Nagumo, Masashi; Kamikado, Shinichi.
Afiliación
  • Mori Y; Department of Anesthesia, Clinical Engineering, National Hospital Organization, Tokyo Medical Center, Tokyo 152-8902.
Masui ; 56(1): 87-90, 2007 Jan.
Article en Ja | MEDLINE | ID: mdl-17243653
ABSTRACT

BACKGROUND:

Retractions of the heart required for exposure and construction of distal anastomoses often decrease R-wave amplitude of ECG and interfere with intraaortic balloon pump (IABP) trigger during off-pump coronary artery bypass (OPCAB). Missing R wave trigger results in asynchronous work of IABP and probably produces hemodynamic instability. We report our early experience with a new interface BPI 202 (Osypka Medical, Inc., USA) for sensing accurate ECG trigger for IABP during OPCAB procedure. CASES Six high-risk patients undergoing multivessel OPCAB using BPI 202 are described.

RESULTS:

With the new interface BPI 202, simulated R wave signal could be processed from an external dual chamber pacemaker sensing surface R wave. The simulated R wave was successfully used for controlling IABP and secured a synchronous work between the heart and IABP during heart retraction maneuver. BPI 202, an interface for IABP appeared to facilitate the intraoperative management of our series of patients.

CONCLUSIONS:

We believe BPI 202 can produce a synchronous work of IABP during OPCAB procedures to high-risk patients and avoid dangerous hemodynamic instability.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Contrapulsación / Puente de Arteria Coronaria Off-Pump / Electrocardiografía / Contrapulsador Intraaórtico / Angina de Pecho Límite: Aged / Humans / Male Idioma: Ja Revista: Masui Año: 2007 Tipo del documento: Article
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Contrapulsación / Puente de Arteria Coronaria Off-Pump / Electrocardiografía / Contrapulsador Intraaórtico / Angina de Pecho Límite: Aged / Humans / Male Idioma: Ja Revista: Masui Año: 2007 Tipo del documento: Article
...