The ideal oxygen/nitrous oxide fresh gas flow sequence with the Anesthesia Delivery Unit machine.
Hendrickx, Jan F A; Cardinael, Sara; Carette, Rik; Lemmens, Hendrikus J M; De Wolf, Andre M.
Afiliación
  • Hendrickx JF; Department of Anesthesiology, Intensive Care and Pain Therapy, Onze Lieve Vrouw Hospital, 9300 Aalst, Belgium. jcnwahendrickx@yahoo.com
J Clin Anesth ; 19(4): 274-9, 2007 Jun.
Article en En | MEDLINE | ID: mdl-17572322
STUDY OBJECTIVE: To determine whether early reduction of oxygen and nitrous oxide fresh gas flow from 6 L/min to 0.7 L/min could be accomplished while maintaining end-expired nitrous oxide concentration > or =50% with an Anesthesia Delivery Unit anesthesia machine. STUDY DESIGN: Prospective, randomized clinical study. SETTING: Large teaching hospital in Belgium. PATIENTS: 53 ASA physical status I and II patients requiring general endotracheal anesthesia and controlled mechanical ventilation. INTERVENTIONS: Patients were randomly assigned to one of 4 groups depending on the duration of high oxygen/nitrous oxide fresh gas flow (two and 4 L/min, respectively) before lowering total fresh gas flow to 0.7 L/min (0.3 and 0.4 L/min oxygen and nitrous oxide, respectively): one, two, three, or 5 minutes (1-minute group, 2-minute group, 3-minute group, and 5-minute group), with n = 10, 12, 13, and 8, respectively. The course of the end-expired nitrous oxide concentration and bellows volume deficit at end-expiration was compared among the 4 groups during the first 30 minutes. RESULTS: At the end of the high-flow period the end-expired nitrous oxide concentration was 35.6 +/- 6.2%, 48.4 +/- 4.8%, 53.7 +/- 8.7%, and 57.3 +/- 1.6% in the 4 groups, respectively. Thereafter, the end-expired nitrous oxide concentration decreased to a nadir of 36.1 +/- 4.5%, 45.4 +/- 3.8%, 50.9 +/- 6.1%, and 55.4 +/- 2.8% after three, 4, 6, and 8 minutes after flows were lowered in the 1- to 5-minute groups, respectively. A decrease in bellows volume was observed in most patients, but was most pronounced in the 2-minute group. The bellows volume deficit gradually faded within 15 to 20 minutes in all 4 groups. CONCLUSIONS: A 3-minute high-flow period (oxygen and nitrous oxide fresh gas flow of 2 and 4 L/min, respectively) suffices to attain and maintain end-expired nitrous oxide concentration > or =50% and ensures an adequate bellows volume during the ensuing low-flow period.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oxígeno / Anestésicos por Inhalación / Anestesiología / Óxido Nitroso Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Anesth Asunto de la revista: ANESTESIOLOGIA Año: 2007 Tipo del documento: Article País de afiliación: Bélgica
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oxígeno / Anestésicos por Inhalación / Anestesiología / Óxido Nitroso Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Anesth Asunto de la revista: ANESTESIOLOGIA Año: 2007 Tipo del documento: Article País de afiliación: Bélgica
...