Your browser doesn't support javascript.
loading
Diamagnetic loop for the first plasma in the KSTAR machine.
Bak, J G; Lee, S G; Ka, E M.
Afiliación
  • Bak JG; National Fusion Research Institute, Gwahangno 113, Yuseong-Gu, Daejeon 305-333, Republic of Korea. jgbak@nfri.re.kr
Rev Sci Instrum ; 79(10): 10F118, 2008 Oct.
Article en En | MEDLINE | ID: mdl-19044602
ABSTRACT
The diamagnetic loop (DL) is installed for the plasma diamagnetic measurement at the first plasma in the Korea superconducting tokamak advanced research (KSTAR) machine. Experimental results from the position measurement of the DL inside the KSTAR vacuum vessel and the vacuum flux measurement by using the DL for the evaluation of the geometrical data and the balance coefficient of the DL for the compensation of the vacuum flux in the diamagnetic measurement are described. In addition, a preliminary work of an instrument for a hardware compensation of the vacuum flux is presented.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2008 Tipo del documento: Article