Chopper system for time resolved experiments with synchrotron radiation.
Rev Sci Instrum ; 80(1): 015101, 2009 Jan.
Article en En | MEDLINE | ID: mdl-19191457
ABSTRACT
A chopper system for time resolved pump-probe experiments with x-ray beams from a synchrotron is described. The system has three parts a water-cooled heatload chopper, a high-speed chopper, and a millisecond shutter. The chopper system, which is installed in beamline ID09B at the European Synchrotron Radiation Facility, provides short x-ray pulses for pump-probe experiments with ultrafast lasers. The chopper system can produce x-ray pulses as short as 200 ns in a continuous beam and repeat at frequencies from 0 to 3 kHz. For bunch filling patterns of the synchrotron with pulse separations greater than 100 ns, the high-speed chopper can isolate single 100 ps x-ray pulses that are used for the highest time resolution. A new rotor in the high-speed chopper is presented with a single pulse (100 ps) and long pulse (10 micros) option. In white beam experiments, the heatload of the (noncooled) high-speed chopper is lowered by a heatload chopper, which absorbs 95% of the incoming power without affecting the pulses selected by the high speed chopper.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Rayos X / Sincrotrones / Rayos Láser Idioma: En Revista: Rev Sci Instrum Año: 2009 Tipo del documento: Article País de afiliación: Francia