Your browser doesn't support javascript.
loading
Comparison of the BD Phoenix system with the cefoxitin disk diffusion test for detection of methicillin resistance in Staphylococcus aureus and coagulase-negative staphylococci.
Mencacci, Antonella; Montecarlo, Ines; Gonfia, Francesca; Moretti, Amedeo; Cardaccia, Angela; Farinelli, Senia; Pagliochini, Maria Rita; Giuliani, Angela; Basileo, Michela; Pasticci, Maria Bruna; Bistoni, Francesco.
Afiliación
  • Mencacci A; Department of Experimental Medicine and Biochemical Sciences, University of Perugia, Perugia, Italy. mencacci@unipg.it
J Clin Microbiol ; 47(7): 2288-91, 2009 Jul.
Article en En | MEDLINE | ID: mdl-19439535
ABSTRACT
The BD Phoenix system was compared to the cefoxitin disk diffusion test for detection of methicillin (meticillin) resistance in 1,066 Staphylococcus aureus and 1,121 coagulase-negative staphylococcus (CoNS) clinical isolates. The sensitivity for Phoenix was 100%. The specificities were 99.86% for S. aureus and 88.4% for CoNS.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Staphylococcus / Cefoxitina / Resistencia a la Meticilina / Antibacterianos Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Microbiol Año: 2009 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Staphylococcus / Cefoxitina / Resistencia a la Meticilina / Antibacterianos Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Microbiol Año: 2009 Tipo del documento: Article País de afiliación: Italia
...