Comparison of the BD Phoenix system with the cefoxitin disk diffusion test for detection of methicillin resistance in Staphylococcus aureus and coagulase-negative staphylococci.
J Clin Microbiol
; 47(7): 2288-91, 2009 Jul.
Article
en En
| MEDLINE
| ID: mdl-19439535
ABSTRACT
The BD Phoenix system was compared to the cefoxitin disk diffusion test for detection of methicillin (meticillin) resistance in 1,066 Staphylococcus aureus and 1,121 coagulase-negative staphylococcus (CoNS) clinical isolates. The sensitivity for Phoenix was 100%. The specificities were 99.86% for S. aureus and 88.4% for CoNS.