Sample evaporation from pierceable cups: still an important source of analytical error.
Schouwers, Sofie; Cuypers, Eric; Vervaet, Sabine; Uyttenbroeck, Wim; Neels, Hugo.
Afiliación
  • Schouwers S; Laboratory for Clinical Chemistry and Toxicology, ZNA Stuivenberg, Antwerp, Belgium.
Clin Biochem ; 43(18): 1464-7, 2010 Dec.
Article en En | MEDLINE | ID: mdl-20850427
BACKGROUND: We illustrate the impact of sample evaporation on analytical results in laboratory practice and highlight preventive measures. METHODS: Plasma (n=10) was analysed for glucose, Na(+), HCO(3)(-) and calcium on six different sample configurations at 5 time points within 2h. RESULTS: With time glucose, Na(+) and calcium values increased and HCO(3)(-) values decreased in a clinically significant way. CONCLUSIONS: Analytical error due to evaporation may be significant, but can be reduced with optimal sample handling. A pierceable cover does not prevent loss of HCO(3)(-).
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Recolección de Muestras de Sangre Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: Clin Biochem Año: 2010 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Recolección de Muestras de Sangre Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: Clin Biochem Año: 2010 Tipo del documento: Article País de afiliación: Bélgica
...