Your browser doesn't support javascript.
loading
Assessment of tissue autofluorescence and reflectance for oral cavity cancer screening.
Sweeny, Larissa; Dean, Nichole R; Magnuson, J Scott; Carroll, William R; Clemons, Lisa; Rosenthal, Eben L.
Afiliación
  • Sweeny L; Department of Surgery, Division of Otolaryngology-Head and Neck Surgery, University of Alabama at Birmingham, Birmingham, Alabama 35233, USA.
Otolaryngol Head Neck Surg ; 145(6): 956-60, 2011 Dec.
Article en En | MEDLINE | ID: mdl-21804026
ABSTRACT

OBJECTIVE:

Although approved by the US Food and Drug Administration for clinical use, the utility of handheld tissue reflectance and autofluorescence devices for screening head and neck cancer patients is poorly defined. There is limited published evidence regarding the efficacy of these devices. The authors investigated the sensitivity and specificity of these modalities compared with standard examination. STUDY

DESIGN:

Prospective, cross-sectional analysis.

SETTING:

Tertiary care medical center. SUBJECTS AND

METHODS:

Patients who were treated previously for head and neck cancer (n = 88) between 2009 and 2010 were included. Patients were screened using white light visualization (standard of care) and compared with tissue reflectance and autofluorescence visualization. Screening results were compared with biopsy or long-term follow-up.

RESULTS:

Autofluorescence visualization had a specificity of 81% and a sensitivity of 50% for detecting oral cavity cancer, whereas white light visualization had a specificity of 98% and a sensitivity of 50%. Tissue reflectance visualization had low sensitivity (0%) and good specificity (86%). The power of this study was insufficient to compare the positive and negative predictive values of standard white light examination (50% and 98%, respectively) to tissue autofluorescence (11% and 97%) or reflectance (0% and 95%). In addition, stratification by previous radiation therapy found no statistically significant difference in screening results.

CONCLUSION:

Standard clinical lighting has a higher specificity than tissue reflectance and autofluorescence visualization for detection of disease in patients with a history of head and neck cancer. This study does not support the added costs associated with these devices.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 2_ODS3 Problema de salud: 2_muertes_prematuras_enfermedades_notrasmisibles Asunto principal: Neoplasias de la Boca / Tamizaje Masivo / Fluorescencia / Microscopía de Interferencia / Recurrencia Local de Neoplasia Tipo de estudio: Diagnostic_studies / Evaluation_studies / Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Otolaryngol Head Neck Surg Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 2_ODS3 Problema de salud: 2_muertes_prematuras_enfermedades_notrasmisibles Asunto principal: Neoplasias de la Boca / Tamizaje Masivo / Fluorescencia / Microscopía de Interferencia / Recurrencia Local de Neoplasia Tipo de estudio: Diagnostic_studies / Evaluation_studies / Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Otolaryngol Head Neck Surg Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos
...