Mix-mode TiO-C18 monolith spin column extraction and GC-MS for the simultaneous assay of organophosphorus compounds and glufosinate, and glyphosate in human serum and urine.
Anal Sci
; 27(10): 999-1005, 2011.
Article
en En
| MEDLINE
| ID: mdl-21985924
ABSTRACT
A rapid, specific, and sensitive method for the simultaneous quantitation of organophosphates (fenitrothion (MEP), malathion, and phenthoate (PAP)), glufosinate (GLUF), and glyphosate (GLYP) in human serum and urine by gas chromatography-mass spectrometry (GC-MS) has been validated. All of the targeted compounds together with the internal standard were extracted from the serum and urine using a mix-mode TiO-C(18) monolithic spin column. The recovery of organophosphates from serum and urine ranged from 12.7 to 49.5%. The recovery of GLUF and GLYP from serum and urine ranged from 1.9 to 7.9%. The intra- and inter-accuracy and precision (expressed as relative standard deviation, %RSD) were within 96.7-107.7% and 4.0-13.8%, respectively. The detection and quantitation limits for serum and urine were 0.1 and 0.1 µg/ml, respectively, for organophosphates, 0.1 and 0.5 µg/ml, respectively for GLUF and GLYP. The method had linear calibration curves ranging from 0.1 to 25.0 µg/ml for organophosphates and 0.5-100.0 µg/ml for GLUF, and GLYP. The validated method was successfully applied to a clinical GLYP poisoning case.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Compuestos Organofosforados
/
Glicina
/
Aminobutiratos
Tipo de estudio:
Diagnostic_studies
Límite:
Humans
Idioma:
En
Revista:
Anal Sci
Año:
2011
Tipo del documento:
Article
País de afiliación:
Japón