Your browser doesn't support javascript.
loading
Clinical application of Ultracontrol®: infusion volume and use with different dialyzers.
Albalate Ramón, M; Pérez García, R; de Sequera Ortiz, P; Alcázar Arroyo, R; Corchete Prats, E; Puerta Carretero, M; Ortega Díaz, M; Mosse, A.
Afiliación
  • Albalate Ramón M; Nephrology, Hospital Infanta Leonor, Avenida Gran Vía del Este 80, Madrid, Spain. malbalater@senefro.org
Nefrologia ; 31(6): 683-9, 2011.
Article en En, Es | MEDLINE | ID: mdl-22130284
ABSTRACT

INTRODUCTION:

Recent studies indicate that the survival benefit with post-dilution on line hemodiafiltration (OL-HDF-post) are achieved if the infusion volume (Vinf) is greater than 20L per session, a goal that is not easily achieved due to hemoconcentration problems. Today we have automated techniques to achieve higher performance minimizing the number of alarms as Ultracontrol® (Ultrac). The objective in the first part of study was to evaluate the UltraC performance (expressed as the filtration fraction (FF) and Vinf) and which problems it presents, and in the second part, to study its performance with four different dialyzers. MATERIAL AND

METHODS:

1st period. Nine patients were transferred to OL-HDF-post with UltraC. The first 3 months on OL-HDF all sessions were recorded and compared with hemodialysis sessions in the previous month. 2nd part 18 patients on chronic OL-HDF-post were dialyzed for a week with each of these dialyser FX1000, FX800, Elisio210H and Polyflux210.

RESULTS:

1st period In 3 patients, problems associated with inappropriate pressures emerged. In 3 patients there were problems associated with inadequate PTM and Psist that resolved changing to pressure control. Mean values were maximum Qb 441 (21) (range 350-490) ml/min, Vinf 26.3 (3.3) l/session, FF 30.6 (2.5)%, KT 59.9 (5) l/session. KT increase of 12% compared to HD. 2nd part Polyflux210 required less UltraC withdrawals than the others. Different PTM or Psist were found and determined the need for removal of the system. The KT was adequate. a) The UltraC system reaches FF of 30% with minimal alarms and Vinf higher than 20 l. b) Structural characteristics of dialysers can limit their use with UltraC although they managed to desirable KT and Vinf in a manual way.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Diálisis Renal / Hemodiafiltración / Fallo Renal Crónico Límite: Aged / Female / Humans / Male / Middle aged Idioma: En / Es Revista: Nefrologia Año: 2011 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Diálisis Renal / Hemodiafiltración / Fallo Renal Crónico Límite: Aged / Female / Humans / Male / Middle aged Idioma: En / Es Revista: Nefrologia Año: 2011 Tipo del documento: Article País de afiliación: España
...