Standardization of a 85Sr solution by three methods.
Appl Radiat Isot
; 70(9): 2222-6, 2012 Sep.
Article
en En
| MEDLINE
| ID: mdl-22425416
A (85)Sr solution was measured in two liquid scintillation (LS) counters. Scintillation pulses from the delayed transitions in the (85)Sr decay were by and large lost as the dead time of the LS-counter was much longer than the lifetime of the (85)Rb excited level. The registration probability of these pulses has been defined and introduced into the counting efficiency calculation. The results of the (85)Sr solution standardization by the CIEMAT/NIST and the TDCR methods relative to the results from the 4π(LS)-γ coincidence and anticoincidence method were 0.997±0.012 and 0.971±0.019, respectively.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Radioisótopos
/
Radiometría
/
Radioisótopos de Estroncio
/
Algoritmos
Idioma:
En
Revista:
Appl Radiat Isot
Asunto de la revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Año:
2012
Tipo del documento:
Article
País de afiliación:
Polonia