Symmetric geometric transfer matrix partial volume correction for PET imaging: principle, validation and robustness.
Phys Med Biol
; 57(21): 7101-16, 2012 Nov 07.
Article
en En
| MEDLINE
| ID: mdl-23051703
Limited spatial resolution of positron emission tomography (PET) often requires partial volume correction (PVC) to improve the accuracy of quantitative PET studies. Conventional region-based PVC methods use co-registered high resolution anatomical images (e.g. computed tomography (CT) or magnetic resonance images) to identify regions of interest. Spill-over between regions is accounted for by calculating regional spread functions (RSFs) in a geometric transfer matrix (GTM) framework. This paper describes a new analytically derived symmetric GTM (sGTM) method that relies on spill-over between RSFs rather than between regions. It is shown that the sGTM is mathematically equivalent to Labbe's method; however it is a region-based method rather than a voxel-based method and it avoids handling large matrices. The sGTM method was validated using two three-dimensional (3D) digital phantoms and one physical phantom. A 3D digital sphere phantom with sphere diameters ranging from 5 to 30 mm and a sphere-to-background uptake ratio of 3-to-1 was used. A 3D digital brain phantom was used with four different anatomical regions and a background region with different activities assigned to each region. A physical sphere phantom with the same geometry and uptake as the digital sphere phantom was manufactured and PET-CT images were acquired. Using these three phantoms, the performance of the sGTM method was assessed against that of the GTM method in terms of accuracy, precision, noise propagation and robustness. The robustness was assessed by applying mis-registration errors and errors in estimates of PET point spread function (PSF). In all three phantoms, the results showed that the sGTM method has accuracy similar to that of the GTM method and within 5%. However, the sGTM method showed better precision and noise propagation than the GTM method, especially for spheres smaller than 13 mm. Moreover, the sGTM method was more robust than the GTM method when mis-registration errors or errors in estimates of PSF occur. The improved robustness was more pronounced for smaller objects. In conclusion, the sGTM method was analytically derived and validated. The noise characteristics and robustness of the sGTM method were better than the conventional GTM method.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Imagenología Tridimensional
/
Tomografía de Emisión de Positrones
Tipo de estudio:
Prognostic_studies
Idioma:
En
Revista:
Phys Med Biol
Año:
2012
Tipo del documento:
Article
País de afiliación:
Canadá