Tracheal intubation with a camera embedded in the tube tip (Vivasight(™) ).
Anaesthesia
; 68(1): 74-8, 2013 Jan.
Article
en En
| MEDLINE
| ID: mdl-23121322
ABSTRACT
We studied tracheal intubation in manikins and patients with a camera embedded in the tip of the tracheal tube (Vivasight(™) ). Four people in two teams and two individuals attempted intubation of a manikin through an i-gel(™) 10 times each. The tracheas of 12 patients with a Mallampati grade of 1 were intubated with a Vivasight tracheal tube through a Berman airway, passed over a Frova(™) introducer. All 60 manikin intubations were successful, taking a mean (SD) time of 1.4 (0.5) s. The fastest intubation was performed in 0.5 s. All 12 participants' tracheas were successfully intubated in a median (IQR [range]) time of 90 (70-120 [50-210]) s. Seven participants complained of a sore throat, comparable with earlier findings for standard laryngoscopy and intubation five mild; one moderate; and one severe. Tracheal intubation with the Vivasight through the i-gel or Berman airway is an alternative to existing techniques, against which it should be compared in randomised controlled trials in human participants. It has potential as a fast airway rescue technique.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Laringoscopios
/
Manejo de la Vía Aérea
/
Intubación Intratraqueal
Tipo de estudio:
Clinical_trials
/
Etiology_studies
Límite:
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Anaesthesia
Año:
2013
Tipo del documento:
Article
País de afiliación: