Night glucose control with MD-Logic artificial pancreas in home setting: a single blind, randomized crossover trial-interim analysis.
Nimri, Revital; Muller, Ido; Atlas, Eran; Miller, Shahar; Kordonouri, Olga; Bratina, Natasa; Tsioli, Christiana; Stefanija, Magdalena A; Danne, Thomas; Battelino, Tadej; Phillip, Moshe.
Afiliación
  • Nimri R; The Jesse Z and Sara Lea Shafer Institute for Endocrinology and Diabetes, National Center for Childhood Diabetes, Schneider Children's Medical Center of Israel, Petah Tikva, Israel.
Pediatr Diabetes ; 15(2): 91-9, 2014 Mar.
Article en En | MEDLINE | ID: mdl-23944875
ABSTRACT

BACKGROUND:

Artificial pancreas (AP) systems have shown an improvement in glucose control and a reduced risk of nocturnal hypoglycemia under controlled conditions but remain to be evaluated under daily-life conditions.

OBJECTIVE:

To assess the feasibility, safety, and efficacy of the MD-Logic AP in controlling nocturnal glucose levels in the patient's home.

METHODS:

Two-arm study, each covering four consecutive nights comparing the MD-Logic AP ('closed-loop' arm) with sensor-augmented pump therapy ('control' arm). Fifteen patients (mean age 19 ± 10.4 yr, A1c 7.5 ± 0.5% or 58 ± 5.9 mmol/mol, diabetes duration 9.9 ± 8.2 yr) were randomly assigned either to 'Group A' (first 'closed-loop', then 'control' arm) or to 'Group B' (vice versa). Investigators were masked to treatment intervention. Primary endpoints were the time spent with glucose levels below 70 mg/dL and the percentage of nights in which the mean overnight glucose levels were within 90-140 mg/dL. Endpoint analyses were based on unmodified sensor glucose readings of the four study nights.

RESULTS:

Time of glucose levels spent below 70 mg/dL was significantly shorter on the closed-loop nights than on control nights, median and interquartile range 3.8 (0, 11.6) and 48.7 (0.6, 67.9) min, respectively; p = 0.0034. The percentage of individual nights in which mean overnight glucose level was within 90-140 mg/dL was 67 (33, 88), and 50 (25, 75), under closed-loop and control nights, respectively, with no statistical difference. Secondary endpoint analyses demonstrated significant improvements in hypoglycemia parameters. No serious adverse events were reported.

CONCLUSION:

This interim analysis demonstrates the feasibility, safety, and efficiency of the MD-Logic AP system in home use, and demonstrates an improvement over sensor-augmented pump therapy. (ClinicalTrials.gov identifier NCT01726829).
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 / 2_ODS3 Problema de salud: 1_doencas_nao_transmissiveis / 2_muertes_prematuras_enfermedades_notrasmisibles Asunto principal: Glucemia / Páncreas Artificial / Diabetes Mellitus Tipo 1 / Servicios de Atención de Salud a Domicilio Tipo de estudio: Clinical_trials Límite: Adolescent / Adult / Child / Female / Humans / Male Idioma: En Revista: Pediatr Diabetes Asunto de la revista: ENDOCRINOLOGIA Año: 2014 Tipo del documento: Article País de afiliación: Israel

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 / 2_ODS3 Problema de salud: 1_doencas_nao_transmissiveis / 2_muertes_prematuras_enfermedades_notrasmisibles Asunto principal: Glucemia / Páncreas Artificial / Diabetes Mellitus Tipo 1 / Servicios de Atención de Salud a Domicilio Tipo de estudio: Clinical_trials Límite: Adolescent / Adult / Child / Female / Humans / Male Idioma: En Revista: Pediatr Diabetes Asunto de la revista: ENDOCRINOLOGIA Año: 2014 Tipo del documento: Article País de afiliación: Israel
...