Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics.
Shao, Yiping; Sun, Xishan; Lan, Kejian A; Bircher, Chad; Lou, Kai; Deng, Zhi.
Afiliación
  • Shao Y; Department of Imaging Physics, The University of Texas M D Anderson Cancer Center, Houston, TX, USA.
Phys Med Biol ; 59(5): 1223-38, 2014 Mar 07.
Article en En | MEDLINE | ID: mdl-24556629
ABSTRACT
In this study, we developed a prototype animal PET by applying several novel technologies to use solid-state photomultiplier (SSPM) arrays to measure the depth of interaction (DOI) and improve imaging performance. Each PET detector has an 8 × 8 array of about 1.9 × 1.9 × 30.0 mm(3) lutetium-yttrium-oxyorthosilicate scintillators, with each end optically connected to an SSPM array (16 channels in a 4 × 4 matrix) through a light guide to enable continuous DOI measurement. Each SSPM has an active area of about 3 × 3 mm(2), and its output is read by a custom-developed application-specific integrated circuit to directly convert analogue signals to digital timing pulses that encode the interaction information. These pulses are transferred to and are decoded by a field-programmable gate array-based time-to-digital convertor for coincident event selection and data acquisition. The independent readout of each SSPM and the parallel signal process can significantly improve the signal-to-noise ratio and enable the use of flexible algorithms for different data processes. The prototype PET consists of two rotating detector panels on a portable gantry with four detectors in each panel to provide 16 mm axial and variable transaxial field-of-view (FOV) sizes. List-mode ordered subset expectation maximization image reconstruction was implemented. The measured mean energy, coincidence timing and DOI resolution for a crystal were about 17.6%, 2.8 ns and 5.6 mm, respectively. The measured transaxial resolutions at the center of the FOV were 2.0 mm and 2.3 mm for images reconstructed with and without DOI, respectively. In addition, the resolutions across the FOV with DOI were substantially better than those without DOI. The quality of PET images of both a hot-rod phantom and mouse acquired with DOI was much higher than that of images obtained without DOI. This study demonstrates that SSPM arrays and advanced readout/processing electronics can be used to develop a practical DOI-measureable PET scanner.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fotometría / Interfaz Usuario-Computador / Imagenología Tridimensional / Tomografía de Emisión de Positrones Tipo de estudio: Diagnostic_studies Límite: Animals Idioma: En Revista: Phys Med Biol Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fotometría / Interfaz Usuario-Computador / Imagenología Tridimensional / Tomografía de Emisión de Positrones Tipo de estudio: Diagnostic_studies Límite: Animals Idioma: En Revista: Phys Med Biol Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos
...