The motional Stark effect polarimeter in the HL-2A tokamak.
Rev Sci Instrum
; 85(5): 053508, 2014 May.
Article
en En
| MEDLINE
| ID: mdl-24880370
A 7-channel motional Stark effect polarimeter based on four polarizers and a spectrometer has been developed in the HL-2A tokamak, which is the first time successful utilizing this kind of polarimeter on a tokamak. The accuracy of the angle can reach ±0.25° in the calibration experiments. Pilot experiments of measuring the magnetic pitch angle have been successfully carried out in the weak motional Stark effect plasma discharge with toroidal magnetic field of ~1.3 T and beam energy of ~25 keV/amu. The pitch angles of magnetic field are obtained for 7 spatial points covering 24 cm along major radius with time resolution of 40 ms; the profiles of safety factor are obtained by combining with the Equilibrium and Reconstruction Fitting Code. The core value of safety factor (q) is less than 1 during the sawtooth oscillation and the position of q = 1 surface is well consistent with the results measured by soft X-ray array.