Image plate characterization and absolute calibration to low kilo-electron-volt electrons.
Busold, S; Philipp, K; Otten, A; Roth, M.
Afiliación
  • Busold S; GSI Helmholtzzentrum für Schwerionenforschung, Planckstraße 1, 64291 Darmstadt, Germany.
  • Philipp K; Institus für Kernphysik, Technische Universität Darmstadt, Schlossgartenstraße 9, 64289 Darmstadt, Germany.
  • Otten A; Institus für Kernphysik, Technische Universität Darmstadt, Schlossgartenstraße 9, 64289 Darmstadt, Germany.
  • Roth M; Institus für Kernphysik, Technische Universität Darmstadt, Schlossgartenstraße 9, 64289 Darmstadt, Germany.
Rev Sci Instrum ; 85(11): 113306, 2014 Nov.
Article en En | MEDLINE | ID: mdl-25430109
ABSTRACT
We report on the characterization of an image plate and its absolute calibration to electrons in the low keV energy range (1-30 keV). In our case, an Agfa MD4.0 without protection layer was used in combination with a Fuji FLA7000 scanner. The calibration data are compared to other published data and a consistent picture of the sensitivity of image plates to electrons is obtained, which suggests a validity of the obtained calibration up to 100 keV.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Rev Sci Instrum Año: 2014 Tipo del documento: Article País de afiliación: Alemania